HemosIL-AcuStar

HemosIL AcuStar HIT-IgG(PF4-H). Source: fda.gov

Performance problems could produce false-negative results, potentially delaying diagnosis of a dangerous reaction to heparin.

The U.S. Food and Drug Administration (FDA) has identified a recall of certain lots of the HemosIL AcuStar HIT-IgG(PF4-H) diagnostic test, manufactured by Instrumentation Laboratories (Werfen), as the most serious type of medical device recall.

The recall involves a performance problem that could cause the test to incorrectly report negative results in patients being evaluated for heparin-induced thrombocytopenia (HIT), a potentially life-threatening reaction to the blood-thinning medication heparin.

Why the Recall Matters

Heparin is commonly used in hospitals to prevent or treat blood clots. In some patients, however, it can trigger an immune reaction that reduces platelet counts and increases the risk of dangerous blood clots.

The HemosIL AcuStar HIT-IgG(PF4-H) test helps clinicians evaluate patients for this condition by detecting antibodies associated with HIT.

According to the FDA, customer complaints identified problems with test calibration and inaccurate results. Werfen subsequently confirmed that certain lots could produce results with a negative bias of 20% to 35%, meaning the test could report antibody levels lower than their actual values.

More concerning, the problem may not be reliably detected through routine calibration and quality control testing.

As a result, some patient samples could be incorrectly reported below the test’s cutoff of 1.00 U/mL and interpreted as negative for HIT antibodies.

A false-negative result could delay recognition of HIT and influence treatment decisions. Continued heparin exposure in a patient with HIT could increase the risk of serious thrombotic complications.

Products Affected

The recall affects four lots of HemosIL AcuStar HIT-IgG(PF4-H):

Part number Affected lot Expiration date
0009800127 B40005 January 16, 2027
0009800127 B38062 November 1, 2026
0009802028 B40006 January 16, 2027
0009802028 B38063 November 1, 2026

Actions for Healthcare Facilities

Werfen has instructed affected customers to identify and destroy remaining products from the recalled lots, document the destruction, and arrange alternative HIT antibody testing methods. The company has also warned that a supply interruption is expected. Werfen’s investigation into the root cause remains ongoing.

Laboratories and treating clinicians should determine whether previously reported patient results require review, follow-up, or repeat testing based on clinical judgment and applicable procedures.

As of September 3, 2026, the company had reported no serious injuries or deaths associated with the issue.

Regulatory Intelligence Perspective

This recall illustrates an important challenge involving diagnostic devices, that a test can produce an apparently valid result even when its analytical performance is compromised. The reported negative bias is particularly concerning because routine calibration and quality control procedures may not reliably identify the problem.

For diagnostic manufacturers and clinical laboratories, the recall underscores the importance of investigating performance complaints, evaluating potential effects on previously reported patient results, and assessing whether existing quality controls are capable of detecting clinically significant failures.

The case also demonstrates why postmarket surveillance and effective corrective action processes are essential throughout a diagnostic device’s commercial life.

Additional information

U.S. customers may contact Werfen Client Services at 1-800-955-9525. FDA recall information is available through the agency’s Medical Device Recalls and Early Alerts website.

Urgent Recall Alert for Methocarbamol Injection by Eugia US LLC 

Attention Consumers, Healthcare Providers, and Distributors:  Eugia US LLC, formerly known as AuroMedics Pharma LLC, announced a voluntary recall on March 28, 2024 of lot number 3MC23011 of Methocarbamol Injection, USP 1000 mg/10 mL (100mg/mL) 10mL Single Dose Vial....

Amneal Recalls Vancomycin Hydrochloride Oral Solution

Some bottles may be super potent Attention Patients, Health Professionals, and Pharmacies: Amneal Pharmaceuticals issued a safety notice on March 27 for a recall of specific lots of Vancomycin Hydrochloride for Oral Solution USP, 250 mg/ 5mL. This recall has been...

FDA Proposes Ban of Certain ESDs

Ban would affect devices used to treat aggressive/self-injurious behavior In a landmark move, the U.S. Food and Drug Administration (FDA) has proposed a ban on electrical stimulation devices (ESDs) designed to reduce or stop self-injurious or aggressive behavior. This...

Camber Recalls Atovaquone Oral Suspension USP

Camber Recalls Atovaquone Oral Suspension USP for Potential Contamination Camber Pharmaceuticals Inc. recently issued a voluntary nationwide recall of Atovaquone Oral Suspension USP due to potential contamination. The recall was announced by the U.S. Food and Drug...

Exposure to toxic materials during hemodialysis and dialysis

FDA announcement The US Food and Drug Administration (FDA) issued an announcement on March 7 to update its May 2022 letter to healthcare providers regarding the potential risks of exposure to toxic compounds when using hemodialysis and peritoneal dialysis. This update...

FDA Explores Dental Device Safety Concerns

Dental Devices to Treat Sleep Apnea and TMD The FDA is currently evaluating safety concerns related to certain dental devices called fixed (non-removable) palatal expanders, which are used on adults for jaw remodeling or to treat conditions like obstructive sleep...

Update – Carcinoma Inside Capsule of Breast Implants

Squamous Cell Carcinoma and Various Lymphomas Located in the Capsule around Breast Implants On March 22, the US Food and Drug Administration (FDA) issued an update to its March 8 safety communication regarding reports of squamous cell carcinoma (SCC) and lymphomas in...

Carcinoma and Lymphomas Found Near Breast Implants

Squamous Cell Carcinoma and Various Lymphomas Located in the Capsule around Breast Implants On March 8, the US Food and Drug Administration (FDA) issued a safety communication concerning reports of squamous cell carcinoma (SCC) and various types of lymphomas in the...

FDA Warning Letter to Olympus Medical Systems

Letter addresses several violations of FDA&C Act On March 15, 2023, the Food and Drug Administration (FDA) issued a Warning Letter to Olympus Medical Systems Corp. The letter addressed several violations of the Federal Food, Drug, and Cosmetic Act (FD&C Act)...

FDA Warning Letter to Jackson Clinic

Warning Letter Sites Violations of Law  In early February of this year, the U.S. Food and Drug Administration (FDA) issued a Warning Letter to Jackson Health & Wellness Clinic. The business was cited for several violations related to its marketing and...

Medical Devices
Medical Device Recalls
Medical Device Warning Letters