Device Recall: ZOLL AutoPulse NXT Resuscitation System
Device may fail to deliver adequate CPRUrgent Class I Medical Device Recall
ZOLL Circulation, Inc. has issued a Class I recall of its AutoPulse NXT Resuscitation System (Model 200, Software Version 2.0.1) due to a system error that may cause the device to stop chest compressions or deliver compressions that are too shallow.
Device Use
The AutoPulse NXT is a battery-powered CPR device used on adults with no pulse or spontaneous breathing. It automates chest compressions and is intended for emergency use when CPR is needed.
What’s the Problem?
A failure code (FC1060) in the AutoPulse NXT prevents the device from correctly measuring compression depth. This could result in interrupted or inadequate CPR, posing serious risks such as:
- Reduced blood flow to vital organs
- Anoxic brain injury (due to lack of oxygen)
- Death
No injuries or deaths have been reported so far.
Affected Devices
- Product: AutoPulse NXT Platform
- Model: 200
- Catalog Number: 8700-001070-1
- Software Version: 2.0.1
A full list of affected devices follows this post.
What You Should Do
Zoll’s March 4, 2025 Urgent Medical Device Recall Notice advises:
- Immediately stop using any affected AutoPulse NXT units
- Return devices to ZOLL upon receipt of a loaner platform
- ZOLL will repair and retest devices by replacing the faulty center pulley assembly
Questions?
Contact ZOLL Circulation, Inc. at 408-541-2140 for more information.
Read the FDA recall warning here.
See recent device recalls here.
Affected Devices
|
UDI |
Serial Number |
| (01) 00849111003384 | 1179 |
| (01) 00849111003384 | 1168 |
| (01) 00849111003384 | 1178 |
| (01) 00849111003384 | 1184 |
| (01) 00849111003384 | 1188 |
| (01) 00849111003384 | 1197 |
| (01) 00849111003384 | 1109 |
| (01) 00849111003384 | 1110 |
| (01) 00849111003384 | 1126 |
| (01) 00849111003384 | 1166 |
| (01) 00849111003384 | 1169 |
| (01) 00849111003384 | 1152 |
| (01) 00849111003384 | 1199 |
| (01) 00849111003384 | 1202 |
| (01) 00849111003384 | 1111 |
| (01) 00849111003384 | 1185 |
| (01) 00849111003384 | 1198 |
| (01) 00849111003384 | 1203 |
| (01) 00849111003384 | 1206 |
| (01) 00849111003384 | 1208 |
| (01) 00849111003384 | 1211 |
| (01) 00849111003384 | 1212 |
| (01) 00849111003384 | 1172 |
| (01) 00849111003384 | 1182 |
| (01) 00849111003384 | 1130 |
| (01) 00849111003384 | 1128 |
| (01) 00849111003384 | 1129 |
| (01) 00849111003384 | 1189 |
| (01) 00849111003384 | 1183 |
| (01) 00849111003384 | 1131 |
| (01) 00849111003384 | 1134 |
| (01) 00849111003384 | 1191 |
| (01) 00849111003384 | 1186 |
| (01) 00849111003384 | 1192 |
| (01) 00849111003384 | 1170 |
| (01) 00849111003384 | 1173 |
| (01) 00849111003384 | 1175 |
| (01) 00849111003384 | 1180 |
| (01) 00849111003384 | 1174 |
| (01) 00849111003384 | 1176 |
| (01) 00849111003384 | 1187 |
| (01) 00849111003384 | 1177 |
| (01) 00849111003384 | 1116 |
| (01) 00849111003384 | 1165 |
| (01) 00849111003384 | 1118 |
| (01) 00849111003384 | 1119 |
| (01) 00849111003384 | 1193 |
| (01) 00849111003384 | 1138 |
| (01) 00849111003384 | 1139 |
| (01) 00849111003384 | 1145 |
| (01) 00849111003384 | 1146 |
| (01) 00849111003384 | 1147 |
| (01) 00849111003384 | 1148 |
| (01) 00849111003384 | 1149 |
| (01) 00849111003384 | 1150 |
| (01) 00849111003384 | 1153 |
| (01) 00849111003384 | 1155 |
| (01) 00849111003384 | 1159 |
| (01) 00849111003384 | 1161 |
| (01) 00849111003384 | 1162 |
| (01) 00849111003384 | 1163 |
| (01) 00849111003384 | 1164 |
| (01) 00849111003384 | 1181 |
| (01) 00849111003384 | 1141 |
| (01) 00849111003384 | 1108 |
| (01) 00849111003384 | 1190 |
| (01) 00849111003384 | 1195 |
| (01) 00849111003384 | 1120 |
| (01) 00849111003384 | 1121 |
| (01) 00849111003384 | 1122 |
| (01) 00849111003384 | 1123 |
| (01) 00849111003384 | 1107 |
| (01) 00849111003384 | 1112 |
| (01) 00849111003384 | 1113 |
| (01) 00849111003384 | 1114 |
| (01) 00849111003384 | 1115 |
| (01) 00849111003384 | 1137 |
| (01) 00849111003384 | 1140 |
| (01) 00849111003384 | 1196 |
Postmarket Surveillance – from QSR to QMSR
How QSR Postmarket Surveillance Will Evolve Under the QMSR As medical device manufacturers transition from the FDA’s Quality System Regulation (QSR) to the new Quality Management System Regulation (QMSR), one area undergoing significant change is postmarket...
Medical Device Labeling
In the world of medical devices, proper labeling isn't just a regulatory requirement, it is a vital component of patient safety, product usability, and market access. Whether you're a manufacturer navigating FDA regulations, a distributor handling multilingual...