Device Recall: ZOLL AutoPulse NXT Resuscitation System
Device may fail to deliver adequate CPRUrgent Class I Medical Device Recall
ZOLL Circulation, Inc. has issued a Class I recall of its AutoPulse NXT Resuscitation System (Model 200, Software Version 2.0.1) due to a system error that may cause the device to stop chest compressions or deliver compressions that are too shallow.
Device Use
The AutoPulse NXT is a battery-powered CPR device used on adults with no pulse or spontaneous breathing. It automates chest compressions and is intended for emergency use when CPR is needed.
What’s the Problem?
A failure code (FC1060) in the AutoPulse NXT prevents the device from correctly measuring compression depth. This could result in interrupted or inadequate CPR, posing serious risks such as:
- Reduced blood flow to vital organs
- Anoxic brain injury (due to lack of oxygen)
- Death
No injuries or deaths have been reported so far.
Affected Devices
- Product: AutoPulse NXT Platform
- Model: 200
- Catalog Number: 8700-001070-1
- Software Version: 2.0.1
A full list of affected devices follows this post.
What You Should Do
Zoll’s March 4, 2025 Urgent Medical Device Recall Notice advises:
- Immediately stop using any affected AutoPulse NXT units
- Return devices to ZOLL upon receipt of a loaner platform
- ZOLL will repair and retest devices by replacing the faulty center pulley assembly
Questions?
Contact ZOLL Circulation, Inc. at 408-541-2140 for more information.
Read the FDA recall warning here.
See recent device recalls here.
Affected Devices
|
UDI |
Serial Number |
| (01) 00849111003384 | 1179 |
| (01) 00849111003384 | 1168 |
| (01) 00849111003384 | 1178 |
| (01) 00849111003384 | 1184 |
| (01) 00849111003384 | 1188 |
| (01) 00849111003384 | 1197 |
| (01) 00849111003384 | 1109 |
| (01) 00849111003384 | 1110 |
| (01) 00849111003384 | 1126 |
| (01) 00849111003384 | 1166 |
| (01) 00849111003384 | 1169 |
| (01) 00849111003384 | 1152 |
| (01) 00849111003384 | 1199 |
| (01) 00849111003384 | 1202 |
| (01) 00849111003384 | 1111 |
| (01) 00849111003384 | 1185 |
| (01) 00849111003384 | 1198 |
| (01) 00849111003384 | 1203 |
| (01) 00849111003384 | 1206 |
| (01) 00849111003384 | 1208 |
| (01) 00849111003384 | 1211 |
| (01) 00849111003384 | 1212 |
| (01) 00849111003384 | 1172 |
| (01) 00849111003384 | 1182 |
| (01) 00849111003384 | 1130 |
| (01) 00849111003384 | 1128 |
| (01) 00849111003384 | 1129 |
| (01) 00849111003384 | 1189 |
| (01) 00849111003384 | 1183 |
| (01) 00849111003384 | 1131 |
| (01) 00849111003384 | 1134 |
| (01) 00849111003384 | 1191 |
| (01) 00849111003384 | 1186 |
| (01) 00849111003384 | 1192 |
| (01) 00849111003384 | 1170 |
| (01) 00849111003384 | 1173 |
| (01) 00849111003384 | 1175 |
| (01) 00849111003384 | 1180 |
| (01) 00849111003384 | 1174 |
| (01) 00849111003384 | 1176 |
| (01) 00849111003384 | 1187 |
| (01) 00849111003384 | 1177 |
| (01) 00849111003384 | 1116 |
| (01) 00849111003384 | 1165 |
| (01) 00849111003384 | 1118 |
| (01) 00849111003384 | 1119 |
| (01) 00849111003384 | 1193 |
| (01) 00849111003384 | 1138 |
| (01) 00849111003384 | 1139 |
| (01) 00849111003384 | 1145 |
| (01) 00849111003384 | 1146 |
| (01) 00849111003384 | 1147 |
| (01) 00849111003384 | 1148 |
| (01) 00849111003384 | 1149 |
| (01) 00849111003384 | 1150 |
| (01) 00849111003384 | 1153 |
| (01) 00849111003384 | 1155 |
| (01) 00849111003384 | 1159 |
| (01) 00849111003384 | 1161 |
| (01) 00849111003384 | 1162 |
| (01) 00849111003384 | 1163 |
| (01) 00849111003384 | 1164 |
| (01) 00849111003384 | 1181 |
| (01) 00849111003384 | 1141 |
| (01) 00849111003384 | 1108 |
| (01) 00849111003384 | 1190 |
| (01) 00849111003384 | 1195 |
| (01) 00849111003384 | 1120 |
| (01) 00849111003384 | 1121 |
| (01) 00849111003384 | 1122 |
| (01) 00849111003384 | 1123 |
| (01) 00849111003384 | 1107 |
| (01) 00849111003384 | 1112 |
| (01) 00849111003384 | 1113 |
| (01) 00849111003384 | 1114 |
| (01) 00849111003384 | 1115 |
| (01) 00849111003384 | 1137 |
| (01) 00849111003384 | 1140 |
| (01) 00849111003384 | 1196 |