FDA Alert: Abiomed Impella Blood Pump Urgent Recall

Controller may fail to detect the Impella device

Abiomed Impella blood pump controllers

Controller May Fail to Detect Device

Abiomed has issued an urgent safety notice for its Automated Impella Controller (AIC) due to a potentially serious issue where the system may fail to detect the Impella blood pump when connected.

This malfunction could result in inadequate cardiac support, which is especially dangerous for patients in cardiogenic shock.

Affected Products, Product Codes, UDI-DIs, and Serial Numbers (click on the link to download a list.)

Key Details

  • Issue: The controller may not recognize the connected pump during startup or transfer between consoles.
  • No visual alarm appears on screen to indicate a problem.
  • Reported Cases: 3 patient deaths have been linked to this issue as of June 13; no serious injuries reported.
  • Impacted Devices: All AICs across the Abiomed Impella pump line.

Recommended Actions

  • Always have a backup AIC available.
  • If the pump isn’t detected within 20 seconds:
    • Switch back to the previous console.
    • Reboot the non-responsive console.
    • Start the case on a different AIC if needed.
  • Distribute this alert to all relevant clinical staff and facilities.
  • Forward notice to any facility that received affected products.

Why It Matters

Failure to detect the pump could lead to delayed or absent hemodynamic support, putting critically ill patients at high risk.

For questions or to report adverse events, contact Abiomed at (978) 646-1400.

The FDA is actively reviewing this high-risk issue and will provide updates here as more information becomes available.

Read more about Mechanical Circulatory Support (MCS) devices

GE HealthCare Anesthesia Systems Recall Due to Ventilation Risk

GE Carestation Anesthesia Systems GE HealthCare has issued a high-priority correction for certain Carestation anesthesia systems due to a potentially life-threatening risk when used in Volume Control Ventilation (VCV) mode. Affected Devices Carestation 620, 650, 650c...

A Dangerous and Unapproved Drug Sold as a “Cure”

The dangers of tianeptine FDA warns that tianeptine is sold online and even at gas stations as a supposed treatment for depression, anxiety, pain, and opioid withdrawal. But don’t be fooled — this drug is not approved by the FDA for any medical use in the U.S. and has...

Formaldehyde in Hair Smoothing Products: Know the Risks

Know the Risks Before You Straighten Thinking of straightening or smoothing your hair? Many salon hair treatments release formaldehyde gas — a known human carcinogen — when heated. This chemical can pose serious short- and long-term health risks to both clients and...