FDA Alert: Abiomed Impella Blood Pump Urgent Recall
Controller may fail to detect the Impella deviceAbiomed Impella blood pump controllers
Controller May Fail to Detect Device
Abiomed has issued an urgent safety notice for its Automated Impella Controller (AIC) due to a potentially serious issue where the system may fail to detect the Impella blood pump when connected.
This malfunction could result in inadequate cardiac support, which is especially dangerous for patients in cardiogenic shock.
Affected Products, Product Codes, UDI-DIs, and Serial Numbers (click on the link to download a list.)
Key Details
- Issue: The controller may not recognize the connected pump during startup or transfer between consoles.
- No visual alarm appears on screen to indicate a problem.
- Reported Cases: 3 patient deaths have been linked to this issue as of June 13; no serious injuries reported.
- Impacted Devices: All AICs across the Abiomed Impella pump line.
Recommended Actions
- Always have a backup AIC available.
- If the pump isn’t detected within 20 seconds:
- Switch back to the previous console.
- Reboot the non-responsive console.
- Start the case on a different AIC if needed.
- Distribute this alert to all relevant clinical staff and facilities.
- Forward notice to any facility that received affected products.
Why It Matters
Failure to detect the pump could lead to delayed or absent hemodynamic support, putting critically ill patients at high risk.
For questions or to report adverse events, contact Abiomed at (978) 646-1400.
The FDA is actively reviewing this high-risk issue and will provide updates here as more information becomes available.
Read more about Mechanical Circulatory Support (MCS) devices
Postmarket Surveillance – from QSR to QMSR
How QSR Postmarket Surveillance Will Evolve Under the QMSR As medical device manufacturers transition from the FDA’s Quality System Regulation (QSR) to the new Quality Management System Regulation (QMSR), one area undergoing significant change is postmarket...
Medical Device Labeling
In the world of medical devices, proper labeling isn't just a regulatory requirement, it is a vital component of patient safety, product usability, and market access. Whether you're a manufacturer navigating FDA regulations, a distributor handling multilingual...
QMSR Clause 8 – Measurement Analysis and Improvement
ISO 13485 Clause 8 Transition Notes The QMSR eliminates FDA-specific structure in favor of ISO 13485:2016 clauses. All above ISO 13485 clauses are now explicit requirements under QMSR. New documentation and risk-based decision-making are emphasized throughout the...