FDA Alert: Abiomed Impella Blood Pump Urgent Recall

Controller may fail to detect the Impella device

Abiomed Impella blood pump controllers

Controller May Fail to Detect Device

Abiomed has issued an urgent safety notice for its Automated Impella Controller (AIC) due to a potentially serious issue where the system may fail to detect the Impella blood pump when connected.

This malfunction could result in inadequate cardiac support, which is especially dangerous for patients in cardiogenic shock.

Affected Products, Product Codes, UDI-DIs, and Serial Numbers (click on the link to download a list.)

Key Details

  • Issue: The controller may not recognize the connected pump during startup or transfer between consoles.
  • No visual alarm appears on screen to indicate a problem.
  • Reported Cases: 3 patient deaths have been linked to this issue as of June 13; no serious injuries reported.
  • Impacted Devices: All AICs across the Abiomed Impella pump line.

Recommended Actions

  • Always have a backup AIC available.
  • If the pump isn’t detected within 20 seconds:
    • Switch back to the previous console.
    • Reboot the non-responsive console.
    • Start the case on a different AIC if needed.
  • Distribute this alert to all relevant clinical staff and facilities.
  • Forward notice to any facility that received affected products.

Why It Matters

Failure to detect the pump could lead to delayed or absent hemodynamic support, putting critically ill patients at high risk.

For questions or to report adverse events, contact Abiomed at (978) 646-1400.

The FDA is actively reviewing this high-risk issue and will provide updates here as more information becomes available.

Read more about Mechanical Circulatory Support (MCS) devices

QMSR Clause 7 – Product Realization

ISO 13485 Clause 7 - Product Realization  Clause 7 of ISO 13485, titled Product Realization, outlines the structured approach medical device manufacturers must take to design, develop, produce, and deliver devices that meet customer needs and regulatory requirements....

QMSR Clause 6 – Resource Management

ISO 13485 Clause 6 - Resource Management  Like other QSR-to-QMSR transition themes, there is a close alignment between 21 CFR 820.20(b), 820.25, and 820.70 ISO 13485 Clause 6 in their subjects of resources, personnel, training and documentation, awareness of work...

The Role of a Regulatory Professional

510(k) Submissions The New QSMR Regulatory Intelligence RIM Systems Regulatory Competencies Regulatory and Quality Role of Regulatory Professionals: Core Responsibilities Regulatory professionals play a critical role in ensuring that medical devices and pharmaceutical...