FDA Alert: Abiomed Impella Blood Pump Urgent Recall

Controller may fail to detect the Impella device

Abiomed Impella blood pump controllers

Controller May Fail to Detect Device

Abiomed has issued an urgent safety notice for its Automated Impella Controller (AIC) due to a potentially serious issue where the system may fail to detect the Impella blood pump when connected.

This malfunction could result in inadequate cardiac support, which is especially dangerous for patients in cardiogenic shock.

Affected Products, Product Codes, UDI-DIs, and Serial Numbers (click on the link to download a list.)

Key Details

  • Issue: The controller may not recognize the connected pump during startup or transfer between consoles.
  • No visual alarm appears on screen to indicate a problem.
  • Reported Cases: 3 patient deaths have been linked to this issue as of June 13; no serious injuries reported.
  • Impacted Devices: All AICs across the Abiomed Impella pump line.

Recommended Actions

  • Always have a backup AIC available.
  • If the pump isn’t detected within 20 seconds:
    • Switch back to the previous console.
    • Reboot the non-responsive console.
    • Start the case on a different AIC if needed.
  • Distribute this alert to all relevant clinical staff and facilities.
  • Forward notice to any facility that received affected products.

Why It Matters

Failure to detect the pump could lead to delayed or absent hemodynamic support, putting critically ill patients at high risk.

For questions or to report adverse events, contact Abiomed at (978) 646-1400.

The FDA is actively reviewing this high-risk issue and will provide updates here as more information becomes available.

Read more about Mechanical Circulatory Support (MCS) devices

Urgent Recall: Dr. Reddy’s Levetiracetam Injection

What You Need to Know Dr. Reddy’s Laboratories has issued an urgent recall for one batch (Lot No: A1540076) of Levetiracetam in 0.75% Sodium Chloride Injection, 1,000 mg/100 mL due to a serious labeling error. The infusion bag is incorrectly labeled as Levetiracetam...

QMSR Management Responsibility

Transitioning from QSR to QMSR – Management Responsibility The QMSR Management Responsibility requirements bring a more structured, documented, and proactive approach compared to QSR §820.20. While many concepts remain similar, organizations must focus on: - Stronger...

QMSR Quality Manual

Why a Quality Manual Matters in the QMSR Transition U.S. Manufacturers Adopting ISO 13485 Will Need a Quality Manual While QSR does not mandate a Quality Manual, the QMSR ISO 13485 compliance does. This means manufacturers aligning with QMSR who do not have a quality...