FDA Alert: Abiomed Impella Blood Pump Urgent Recall
Controller may fail to detect the Impella deviceAbiomed Impella blood pump controllers
Controller May Fail to Detect Device
Abiomed has issued an urgent safety notice for its Automated Impella Controller (AIC) due to a potentially serious issue where the system may fail to detect the Impella blood pump when connected.
This malfunction could result in inadequate cardiac support, which is especially dangerous for patients in cardiogenic shock.
Affected Products, Product Codes, UDI-DIs, and Serial Numbers (click on the link to download a list.)
Key Details
- Issue: The controller may not recognize the connected pump during startup or transfer between consoles.
- No visual alarm appears on screen to indicate a problem.
- Reported Cases: 3 patient deaths have been linked to this issue as of June 13; no serious injuries reported.
- Impacted Devices: All AICs across the Abiomed Impella pump line.
Recommended Actions
- Always have a backup AIC available.
- If the pump isn’t detected within 20 seconds:
- Switch back to the previous console.
- Reboot the non-responsive console.
- Start the case on a different AIC if needed.
- Distribute this alert to all relevant clinical staff and facilities.
- Forward notice to any facility that received affected products.
Why It Matters
Failure to detect the pump could lead to delayed or absent hemodynamic support, putting critically ill patients at high risk.
For questions or to report adverse events, contact Abiomed at (978) 646-1400.
The FDA is actively reviewing this high-risk issue and will provide updates here as more information becomes available.
Read more about Mechanical Circulatory Support (MCS) devices
QMSR Frequently Asked Questions
Although the FDA has done an excellent job of laying out QMSR requirements and explaining differences between the QSR and the QMSR, there are questions related to the implementation of the new Quality Management System Regulation (QMSR) that need an answer. This page...
EU MDR Compliance Consulting Services
Medical Devices and Pharma offers comprehensive services to help your business meet the stringent requirements of the EU Medical Device Regulation (MDR). Our expertise ensures that your products are compliant, safe, and ready for the European market.At Medical Devices...
Extensions for EU MDR and IVDR Implementation
Initial EU MDR and IVDR transition periods extended In 2023, the European Union granted an extension for the implementation of the Medical Device Regulation (MDR) and the In Vitro Diagnostic Medical Device Regulation (IVDR) due to various challenges faced by...