FDA Alert: Abiomed Impella Blood Pump Urgent Recall

Controller may fail to detect the Impella device

Abiomed Impella blood pump controllers

Controller May Fail to Detect Device

Abiomed has issued an urgent safety notice for its Automated Impella Controller (AIC) due to a potentially serious issue where the system may fail to detect the Impella blood pump when connected.

This malfunction could result in inadequate cardiac support, which is especially dangerous for patients in cardiogenic shock.

Affected Products, Product Codes, UDI-DIs, and Serial Numbers (click on the link to download a list.)

Key Details

  • Issue: The controller may not recognize the connected pump during startup or transfer between consoles.
  • No visual alarm appears on screen to indicate a problem.
  • Reported Cases: 3 patient deaths have been linked to this issue as of June 13; no serious injuries reported.
  • Impacted Devices: All AICs across the Abiomed Impella pump line.

Recommended Actions

  • Always have a backup AIC available.
  • If the pump isn’t detected within 20 seconds:
    • Switch back to the previous console.
    • Reboot the non-responsive console.
    • Start the case on a different AIC if needed.
  • Distribute this alert to all relevant clinical staff and facilities.
  • Forward notice to any facility that received affected products.

Why It Matters

Failure to detect the pump could lead to delayed or absent hemodynamic support, putting critically ill patients at high risk.

For questions or to report adverse events, contact Abiomed at (978) 646-1400.

The FDA is actively reviewing this high-risk issue and will provide updates here as more information becomes available.

Read more about Mechanical Circulatory Support (MCS) devices

Important Recall Notice for BioMérieux VITEK 2 AST Kit 

Class I Recall Notice   BioMérieux Inc. has issued a crucial recall notice for its VITEK 2 AST kit, a vital tool in determining bacterial susceptibility to antibiotics. The Food and Drug Administration (FDA) classifies this recall as Class I, indicating a severe risk...

SonarMed Inc Recalls Airway Acoustic Sensors

Urgent Class I Recall Notice   SonarMed Inc. has issued a Class I recall for its Airway acoustic sensors due to a critical issue that could lead to serious health consequences for patients. The U.S. Food and Drug Administration (FDA) has identified this as the most...

The IMDRF: International Regulators

The International Medical Device Regulators Forum (IMDRF) plays a pivotal role in shaping global regulatory approaches and standards for medical devices, enhancing safety and innovation across international borders.The International Medical Device Regulators Forum...