FDA Alert: Abiomed Impella Blood Pump Urgent Recall
Controller may fail to detect the Impella deviceAbiomed Impella blood pump controllers
Controller May Fail to Detect Device
Abiomed has issued an urgent safety notice for its Automated Impella Controller (AIC) due to a potentially serious issue where the system may fail to detect the Impella blood pump when connected.
This malfunction could result in inadequate cardiac support, which is especially dangerous for patients in cardiogenic shock.
Affected Products, Product Codes, UDI-DIs, and Serial Numbers (click on the link to download a list.)
Key Details
- Issue: The controller may not recognize the connected pump during startup or transfer between consoles.
- No visual alarm appears on screen to indicate a problem.
- Reported Cases: 3 patient deaths have been linked to this issue as of June 13; no serious injuries reported.
- Impacted Devices: All AICs across the Abiomed Impella pump line.
Recommended Actions
- Always have a backup AIC available.
- If the pump isn’t detected within 20 seconds:
- Switch back to the previous console.
- Reboot the non-responsive console.
- Start the case on a different AIC if needed.
- Distribute this alert to all relevant clinical staff and facilities.
- Forward notice to any facility that received affected products.
Why It Matters
Failure to detect the pump could lead to delayed or absent hemodynamic support, putting critically ill patients at high risk.
For questions or to report adverse events, contact Abiomed at (978) 646-1400.
The FDA is actively reviewing this high-risk issue and will provide updates here as more information becomes available.
Read more about Mechanical Circulatory Support (MCS) devices
Overview of ISO 13485
ISO 13485, Medical devices quality management systems ISO 13485, Medical devices quality management systems, is an international standard that specifies requirements for a quality management system (QMS) for medical device manufacturers. It also outlines requirements...
FDA Quality System Regulation – 21 CFR 820
QMSR CentralFDA Quality System Regulation – 21 CFR 820 21 CFR 820, also known as the Quality System Regulation (QSR), outlines the requirements for the establishment and maintenance of a quality management system for medical device manufacturers in the United States....
Edwards EVOQUE Tricuspid Valve Replacement System
Edwards EVOQUE Tricuspid Valve Replacement System In a groundbreaking move, the FDA has recently approved the Edwards EVOQUE Tricuspid Valve Replacement System for market distribution. The device was developed by Edwards Lifesciences LLC. This innovative system...