FDA Alert: Abiomed Impella Blood Pump Urgent Recall

Controller may fail to detect the Impella device

Abiomed Impella blood pump controllers

Controller May Fail to Detect Device

Abiomed has issued an urgent safety notice for its Automated Impella Controller (AIC) due to a potentially serious issue where the system may fail to detect the Impella blood pump when connected.

This malfunction could result in inadequate cardiac support, which is especially dangerous for patients in cardiogenic shock.

Affected Products, Product Codes, UDI-DIs, and Serial Numbers (click on the link to download a list.)

Key Details

  • Issue: The controller may not recognize the connected pump during startup or transfer between consoles.
  • No visual alarm appears on screen to indicate a problem.
  • Reported Cases: 3 patient deaths have been linked to this issue as of June 13; no serious injuries reported.
  • Impacted Devices: All AICs across the Abiomed Impella pump line.

Recommended Actions

  • Always have a backup AIC available.
  • If the pump isn’t detected within 20 seconds:
    • Switch back to the previous console.
    • Reboot the non-responsive console.
    • Start the case on a different AIC if needed.
  • Distribute this alert to all relevant clinical staff and facilities.
  • Forward notice to any facility that received affected products.

Why It Matters

Failure to detect the pump could lead to delayed or absent hemodynamic support, putting critically ill patients at high risk.

For questions or to report adverse events, contact Abiomed at (978) 646-1400.

The FDA is actively reviewing this high-risk issue and will provide updates here as more information becomes available.

Read more about Mechanical Circulatory Support (MCS) devices

FDA – Do Not Use Non-piercing Devices to Measure Blood Glucose

FDA Safety Communication   On February 21, 2024 the U.S. Food and Drug Administration (FDA) issued a safety communication regarding the use of smartwatches or smart rings that claim to measure blood glucose levels without the need for skin piercing.   Its cautionary...

Medos International Recalls CEREBASE DA Guide Sheath 

Urgent Class I Recall Notice   Medos International Sàrl has issued an urgent recall for the Cerenovus CEREBASE DA Guide Sheath, a neurovascular catheter commonly used in procedures requiring precise navigation and access to blood vessels in the brain. The recall was...

Asensus Surgical Inc Issues Recall for Senhance Surgical System 

Class I Recall Notice   Asensus Surgical Inc. has announced a recall for its Senhance Surgical System due to malfunctions leading to unintended movement of the robotically-assisted surgical device.  Device Use   The Senhance Surgical System is a vital tool utilized in...