Controller may fail to detect the Impella deviceAbiomed Impella blood pump controllers
Controller May Fail to Detect Device
Abiomed has issued an urgent safety notice for its Automated Impella Controller (AIC) due to a potentially serious issue where the system may fail to detect the Impella blood pump when connected.
This malfunction could result in inadequate cardiac support, which is especially dangerous for patients in cardiogenic shock.
Affected Products, Product Codes, UDI-DIs, and Serial Numbers (click on the link to download a list.)
Key Details
- Issue: The controller may not recognize the connected pump during startup or transfer between consoles.
- No visual alarm appears on screen to indicate a problem.
- Reported Cases: 3 patient deaths have been linked to this issue as of June 13; no serious injuries reported.
- Impacted Devices: All AICs across the Abiomed Impella pump line.
Recommended Actions
- Always have a backup AIC available.
- If the pump isn’t detected within 20 seconds:
- Switch back to the previous console.
- Reboot the non-responsive console.
- Start the case on a different AIC if needed.
- Distribute this alert to all relevant clinical staff and facilities.
- Forward notice to any facility that received affected products.
Why It Matters
Failure to detect the pump could lead to delayed or absent hemodynamic support, putting critically ill patients at high risk.
For questions or to report adverse events, contact Abiomed at (978) 646-1400.
The FDA is actively reviewing this high-risk issue and will provide updates here as more information becomes available.
Read more about Mechanical Circulatory Support (MCS) devices
Electrical Stimulation Devices – ESDs
Devices use controlled electrical pulses Electrical stimulation medical devices are sophisticated tools designed to deliver controlled electrical impulses to specific areas of the body for therapeutic purposes. These devices utilize various forms of electrical...
FDA Proposes Ban of Certain ESDs
Ban would affect devices used to treat aggressive/self-injurious behavior In a landmark move, the U.S. Food and Drug Administration (FDA) has proposed a ban on electrical stimulation devices (ESDs) designed to reduce or stop self-injurious or aggressive behavior. This...
FDA Approves Thoracic Stent Graft System
FDA Approves RelayPro Thoracic Stent Graft System for Thoracic Aortic Aneurysms and Penetrating Atherosclerotic Ulcers Treatment The United States Food and Drug Administration (FDA) on March 7, 2023 approved the RelayPro Thoracic Stent Graft System, a new medical...