FDA Alert: Abiomed Impella Blood Pump Urgent Recall

Controller may fail to detect the Impella device

Abiomed Impella blood pump controllers

Controller May Fail to Detect Device

Abiomed has issued an urgent safety notice for its Automated Impella Controller (AIC) due to a potentially serious issue where the system may fail to detect the Impella blood pump when connected.

This malfunction could result in inadequate cardiac support, which is especially dangerous for patients in cardiogenic shock.

Affected Products, Product Codes, UDI-DIs, and Serial Numbers (click on the link to download a list.)

Key Details

  • Issue: The controller may not recognize the connected pump during startup or transfer between consoles.
  • No visual alarm appears on screen to indicate a problem.
  • Reported Cases: 3 patient deaths have been linked to this issue as of June 13; no serious injuries reported.
  • Impacted Devices: All AICs across the Abiomed Impella pump line.

Recommended Actions

  • Always have a backup AIC available.
  • If the pump isn’t detected within 20 seconds:
    • Switch back to the previous console.
    • Reboot the non-responsive console.
    • Start the case on a different AIC if needed.
  • Distribute this alert to all relevant clinical staff and facilities.
  • Forward notice to any facility that received affected products.

Why It Matters

Failure to detect the pump could lead to delayed or absent hemodynamic support, putting critically ill patients at high risk.

For questions or to report adverse events, contact Abiomed at (978) 646-1400.

The FDA is actively reviewing this high-risk issue and will provide updates here as more information becomes available.

Read more about Mechanical Circulatory Support (MCS) devices

FDA Issues Letter to Warn of Abbott Trifecta Valve Risks 

Letter outlines risks associated with the heart valve  On February 27, the FDA released a letter to healthcare providers concerning potential risks associated with Abbott Trifecta valves. The Trifecta valve is a type of heart valve that is implanted in patients during...

FDA Announces Recall of GE Nuclear Medicine Systems 

1200-pound Devices Could Fall on Patients On February 15, the FDA issued a statement announcing a recall of certain nuclear medicine systems manufactured by GE Healthcare. The recall involves the 600/800 Series nuclear medicine systems, which are used to diagnose and...

FDA announces recall of LivaNova LifeSPARC System 

The US Food and Drug Administration (FDA) has announced a recall of the Livanova TandemLife LifeSparc System, a medical device used for circulatory support during heart surgery. The recall was initiated due to the potential for the device to experience unintentional...