FDA Alert: Abiomed Impella Blood Pump Urgent Recall

Controller may fail to detect the Impella device

Abiomed Impella blood pump controllers

Controller May Fail to Detect Device

Abiomed has issued an urgent safety notice for its Automated Impella Controller (AIC) due to a potentially serious issue where the system may fail to detect the Impella blood pump when connected.

This malfunction could result in inadequate cardiac support, which is especially dangerous for patients in cardiogenic shock.

Affected Products, Product Codes, UDI-DIs, and Serial Numbers (click on the link to download a list.)

Key Details

  • Issue: The controller may not recognize the connected pump during startup or transfer between consoles.
  • No visual alarm appears on screen to indicate a problem.
  • Reported Cases: 3 patient deaths have been linked to this issue as of June 13; no serious injuries reported.
  • Impacted Devices: All AICs across the Abiomed Impella pump line.

Recommended Actions

  • Always have a backup AIC available.
  • If the pump isn’t detected within 20 seconds:
    • Switch back to the previous console.
    • Reboot the non-responsive console.
    • Start the case on a different AIC if needed.
  • Distribute this alert to all relevant clinical staff and facilities.
  • Forward notice to any facility that received affected products.

Why It Matters

Failure to detect the pump could lead to delayed or absent hemodynamic support, putting critically ill patients at high risk.

For questions or to report adverse events, contact Abiomed at (978) 646-1400.

The FDA is actively reviewing this high-risk issue and will provide updates here as more information becomes available.

Read more about Mechanical Circulatory Support (MCS) devices

Hospira Recalls Antibiotic for Presence of Glass Particles

Hospira, Inc., a division of Pfizer, announced a Class I recall of its Vancomycin Injection product due to visible glass particles that were identified in one vial of the drug. The recall involves lot 33045BA of Vancomycin Hydrochloride Injection USP 1.5g single dose...

Three Companies Recall Blood Pressure Medication

All recalled products had unacceptable levels of nitrosamines Three pharmaceutical manufacturers have recalled batches of the blood pressure drug quinapril this year for exceeding safe levels of nitrosamines. Luprin Pharmaceuticals, Aurobindo Pharma USA, and Pfizer...

FDA Announces Online Electronic 510(k) Submissions

The U.S. Food and Drug Administration (FDA) announced the issuance of its final guidance document for its template for electronic 510(k) submissions on September 22, 2022. The electronic Submission Template and Resource (eSTAR) electronic submissions template is a...