FDA Alert: Abiomed Impella Blood Pump Urgent Recall
Controller may fail to detect the Impella deviceAbiomed Impella blood pump controllers
Controller May Fail to Detect Device
Abiomed has issued an urgent safety notice for its Automated Impella Controller (AIC) due to a potentially serious issue where the system may fail to detect the Impella blood pump when connected.
This malfunction could result in inadequate cardiac support, which is especially dangerous for patients in cardiogenic shock.
Affected Products, Product Codes, UDI-DIs, and Serial Numbers (click on the link to download a list.)
Key Details
- Issue: The controller may not recognize the connected pump during startup or transfer between consoles.
- No visual alarm appears on screen to indicate a problem.
- Reported Cases: 3 patient deaths have been linked to this issue as of June 13; no serious injuries reported.
- Impacted Devices: All AICs across the Abiomed Impella pump line.
Recommended Actions
- Always have a backup AIC available.
- If the pump isn’t detected within 20 seconds:
- Switch back to the previous console.
- Reboot the non-responsive console.
- Start the case on a different AIC if needed.
- Distribute this alert to all relevant clinical staff and facilities.
- Forward notice to any facility that received affected products.
Why It Matters
Failure to detect the pump could lead to delayed or absent hemodynamic support, putting critically ill patients at high risk.
For questions or to report adverse events, contact Abiomed at (978) 646-1400.
The FDA is actively reviewing this high-risk issue and will provide updates here as more information becomes available.
Read more about Mechanical Circulatory Support (MCS) devices
When a “Good Enough” Design May Fail the User
Advancing from Basic Design to Usability In a world shifting toward user-centered regulation, even the simplest medical devices can demonstrate why human factors and usability engineering matter. The FDA’s Quality Management System Regulation (QMSR) – which aligns...
Why You Should Start Your QMSR Transition Now
Start Preparing for the QMSR – Now The FDA’s Quality Management System Regulation (QMSR) will officially replace the long-standing Quality System Regulation (QSR) in February 2026. For many manufacturers, that may sound like plenty of time. But here’s the truth: You...
Ethicon Issues Safety Alert for Surgical Stapler Cartridge
Endopath Echelon Vascular White Reload for Advanced Placement TipSource: FDAUrgent Class I Medical Device Recall Ethicon Endo-Surgery, LLC has issued an urgent correction for its Endopath Echelon Vascular White Reload for Advanced Placement Tip (35 mm, 4-row). This...