GE HealthCare Anesthesia Systems Recall
potentially life-threatening risk in Volume Control Ventilation (VCV) modeGE Carestation Anesthesia Systems
GE HealthCare has issued a high-priority correction for certain Carestation anesthesia systems due to a potentially life-threatening risk when used in Volume Control Ventilation (VCV) mode.
Affected Devices
- Carestation 620, 650, 650c
- Carestation 750, 750c
These systems are used in hospitals and surgical settings for delivering anesthesia and ventilatory support to patients of all ages.
The Problem
In VCV mode, the devices affected may fail to ventilate properly. If this happens:
- Alarms and warnings will be triggered (e.g., bellows stop moving, “Unable to Drive Bellows” message, apnea alerts).
- If uncorrected, this can lead to serious injury or death from hypoxia (low oxygen levels).
What to Do
GE HealthCare recommends:
- Do NOT use VCV mode until the device is corrected.
- Use only:
- Pressure Control Ventilation (PCV)
- Pressure Control Ventilation Volume Guarantee (PCV-VG)
- Or manual ventilation
- Perform a Ventilation Screening Test on each system.
- Train your staff and ensure they are aware of the risk and temporary workaround.
- Return the Acknowledgement Form provided in the correction notice.
Questions?
Call GE HealthCare Service at 1-800-437-1171.
If your facility uses these systems, take immediate action to prevent harm. GE HealthCare is currently correcting the devices — no need to remove them from service if alternate modes are used safely.
FDA: Do Not Use OTC Products to Remove Moles or Skin Tags
FDA: Do Not Use OTC Products to Remove Moles or Skin Tags Thinking about removing a mole, skin tag, or other skin lesion at home? The FDA strongly advises against it. There are no FDA-approved prescription or over-the-counter (OTC) products for removing moles,...
FDA Alert: Baxter Recalls Spectrum Infusion Pumps
Baxter Spectrum Infusion Pump Alert Date: July 2025Device: Sigma Spectrum Infusion System (V6 and V8 Platforms)Manufacturer: Baxter HealthcareIssue Type: Urgent Medical Device RecallFDA Status: High-risk; active monitoring in progress The FDA has issued an early alert...
How Unsafe Medical Devices Can Slip Through the Cracks
Healthcare professionals and the device industry share responsibility for timely adverse event reportingGlobal device market withdrawals not always followed in US In a world of global trade and instant communication, you might assume that a medical device deemed...