GE HealthCare Anesthesia Systems Recall
potentially life-threatening risk in Volume Control Ventilation (VCV) modeGE Carestation Anesthesia Systems
GE HealthCare has issued a high-priority correction for certain Carestation anesthesia systems due to a potentially life-threatening risk when used in Volume Control Ventilation (VCV) mode.
Affected Devices
- Carestation 620, 650, 650c
- Carestation 750, 750c
These systems are used in hospitals and surgical settings for delivering anesthesia and ventilatory support to patients of all ages.
The Problem
In VCV mode, the devices affected may fail to ventilate properly. If this happens:
- Alarms and warnings will be triggered (e.g., bellows stop moving, “Unable to Drive Bellows” message, apnea alerts).
- If uncorrected, this can lead to serious injury or death from hypoxia (low oxygen levels).
What to Do
GE HealthCare recommends:
- Do NOT use VCV mode until the device is corrected.
- Use only:
- Pressure Control Ventilation (PCV)
- Pressure Control Ventilation Volume Guarantee (PCV-VG)
- Or manual ventilation
- Perform a Ventilation Screening Test on each system.
- Train your staff and ensure they are aware of the risk and temporary workaround.
- Return the Acknowledgement Form provided in the correction notice.
Questions?
Call GE HealthCare Service at 1-800-437-1171.
If your facility uses these systems, take immediate action to prevent harm. GE HealthCare is currently correcting the devices — no need to remove them from service if alternate modes are used safely.
Urgent Recall: Dr. Reddy’s Levetiracetam Injection
What You Need to Know Dr. Reddy’s Laboratories has issued an urgent recall for one batch (Lot No: A1540076) of Levetiracetam in 0.75% Sodium Chloride Injection, 1,000 mg/100 mL due to a serious labeling error. The infusion bag is incorrectly labeled as Levetiracetam...
QMSR Management Responsibility
Transitioning from QSR to QMSR – Management Responsibility The QMSR Management Responsibility requirements bring a more structured, documented, and proactive approach compared to QSR §820.20. While many concepts remain similar, organizations must focus on: - Stronger...
QMSR Quality Manual
Why a Quality Manual Matters in the QMSR Transition U.S. Manufacturers Adopting ISO 13485 Will Need a Quality Manual While QSR does not mandate a Quality Manual, the QMSR ISO 13485 compliance does. This means manufacturers aligning with QMSR who do not have a quality...