GE HealthCare Anesthesia Systems Recall

potentially life-threatening risk in Volume Control Ventilation (VCV) mode

GE Carestation Anesthesia Systems

GE HealthCare has issued a high-priority correction for certain Carestation anesthesia systems due to a potentially life-threatening risk when used in Volume Control Ventilation (VCV) mode.

Affected Devices

  • Carestation 620, 650, 650c
  • Carestation 750, 750c

These systems are used in hospitals and surgical settings for delivering anesthesia and ventilatory support to patients of all ages.

The Problem

In VCV mode, the devices affected may fail to ventilate properly. If this happens:

  • Alarms and warnings will be triggered (e.g., bellows stop moving, “Unable to Drive Bellows” message, apnea alerts).
  • If uncorrected, this can lead to serious injury or death from hypoxia (low oxygen levels).

What to Do

GE HealthCare recommends:

  1. Do NOT use VCV mode until the device is corrected.
  2. Use only:
    • Pressure Control Ventilation (PCV)
    • Pressure Control Ventilation Volume Guarantee (PCV-VG)
    • Or manual ventilation
  3. Perform a Ventilation Screening Test on each system.
  4. Train your staff and ensure they are aware of the risk and temporary workaround.
  5. Return the Acknowledgement Form provided in the correction notice.

Questions?

Call GE HealthCare Service at 1-800-437-1171.

If your facility uses these systems, take immediate action to prevent harm. GE HealthCare is currently correcting the devices — no need to remove them from service if alternate modes are used safely.

QMSR Design and Development Controls

From Design Control to Design and Development Both the QSR and QMSR have the same design and development phases of planning, inputs, outputs, verification, validation, transfer, and design changes. All of the 21 CFR 820.30 Design Control requirements are covered in...

ISO 9000 Clause 3 Definitions

Establishing a Common Language for Quality Management Systems ISO 9000:2015, Quality management systems — Fundamentals and vocabulary, serves as the cornerstone of the ISO 9000 family of standards, providing essential terminology and principles for quality management...

Request QMSR Information

QMSR Information Request Form Are you ready for the FDA’s transition from the Quality System Regulation (QSR) to the Quality Management System Regulation (QMSR)? The new QMSR framework aligns more closely with ISO 13485, bringing significant changes to how medical...