GE HealthCare Anesthesia Systems Recall
potentially life-threatening risk in Volume Control Ventilation (VCV) modeGE Carestation Anesthesia Systems
GE HealthCare has issued a high-priority correction for certain Carestation anesthesia systems due to a potentially life-threatening risk when used in Volume Control Ventilation (VCV) mode.
Affected Devices
- Carestation 620, 650, 650c
- Carestation 750, 750c
These systems are used in hospitals and surgical settings for delivering anesthesia and ventilatory support to patients of all ages.
The Problem
In VCV mode, the devices affected may fail to ventilate properly. If this happens:
- Alarms and warnings will be triggered (e.g., bellows stop moving, “Unable to Drive Bellows” message, apnea alerts).
- If uncorrected, this can lead to serious injury or death from hypoxia (low oxygen levels).
What to Do
GE HealthCare recommends:
- Do NOT use VCV mode until the device is corrected.
- Use only:
- Pressure Control Ventilation (PCV)
- Pressure Control Ventilation Volume Guarantee (PCV-VG)
- Or manual ventilation
- Perform a Ventilation Screening Test on each system.
- Train your staff and ensure they are aware of the risk and temporary workaround.
- Return the Acknowledgement Form provided in the correction notice.
Questions?
Call GE HealthCare Service at 1-800-437-1171.
If your facility uses these systems, take immediate action to prevent harm. GE HealthCare is currently correcting the devices — no need to remove them from service if alternate modes are used safely.
Risks with Exactech Equinoxe Shoulder System
FDA Safety Communication April 19 Update: Exactech has initiated a recall to remove Equinoxe Shoulder System devices packaged in the defective bags from facilities’ unused inventory. To see if your device is affected by the recall, visit Exactech’s Recall...
Important Recall Notice for BioMérieux VITEK 2 AST Kit
Class I Recall Notice BioMérieux Inc. has issued a crucial recall notice for its VITEK 2 AST kit, a vital tool in determining bacterial susceptibility to antibiotics. The Food and Drug Administration (FDA) classifies this recall as Class I, indicating a severe risk...
SonarMed Inc Recalls Airway Acoustic Sensors
Urgent Class I Recall Notice SonarMed Inc. has issued a Class I recall for its Airway acoustic sensors due to a critical issue that could lead to serious health consequences for patients. The U.S. Food and Drug Administration (FDA) has identified this as the most...