GE HealthCare Anesthesia Systems Recall

potentially life-threatening risk in Volume Control Ventilation (VCV) mode

GE Carestation Anesthesia Systems

GE HealthCare has issued a high-priority correction for certain Carestation anesthesia systems due to a potentially life-threatening risk when used in Volume Control Ventilation (VCV) mode.

Affected Devices

  • Carestation 620, 650, 650c
  • Carestation 750, 750c

These systems are used in hospitals and surgical settings for delivering anesthesia and ventilatory support to patients of all ages.

The Problem

In VCV mode, the devices affected may fail to ventilate properly. If this happens:

  • Alarms and warnings will be triggered (e.g., bellows stop moving, “Unable to Drive Bellows” message, apnea alerts).
  • If uncorrected, this can lead to serious injury or death from hypoxia (low oxygen levels).

What to Do

GE HealthCare recommends:

  1. Do NOT use VCV mode until the device is corrected.
  2. Use only:
    • Pressure Control Ventilation (PCV)
    • Pressure Control Ventilation Volume Guarantee (PCV-VG)
    • Or manual ventilation
  3. Perform a Ventilation Screening Test on each system.
  4. Train your staff and ensure they are aware of the risk and temporary workaround.
  5. Return the Acknowledgement Form provided in the correction notice.

Questions?

Call GE HealthCare Service at 1-800-437-1171.

If your facility uses these systems, take immediate action to prevent harm. GE HealthCare is currently correcting the devices — no need to remove them from service if alternate modes are used safely.

Class I Recall for Stay-Safe Catheter Extension Sets and Adapters 

Urgent Class I Recall Notice   FDA has issued a Class I recall – the most serious type – for certain medical devices manufactured by Fresenius Medical Care. The recall concerns Stay-Safe Catheter Extension Sets and Stay-Safe/Luer Lock Adapters, pivotal components used...

Regulatory Information Management Systems (RIMS)

What Is a RIMS System?  A Regulatory Information Management (RIM) system is a software solution designed to streamline and automate the management of regulatory information and processes within the life sciences industry, including pharmaceuticals, biotechnology, and...

Abbott/Thoratec Recalls HeartMate II and HeartMate 3 LVAS

Recall Alert: HeartMate II and HeartMate 3 Left Ventricular Assist Systems (LVAS)  The FDA has issued a Class I recall, the most serious type, for the HeartMate II and HeartMate 3 Left Ventricular Assist Systems (LVAS) due to potential serious injuries or death...