GE HealthCare Anesthesia Systems Recall
potentially life-threatening risk in Volume Control Ventilation (VCV) modeGE Carestation Anesthesia Systems
GE HealthCare has issued a high-priority correction for certain Carestation anesthesia systems due to a potentially life-threatening risk when used in Volume Control Ventilation (VCV) mode.
Affected Devices
- Carestation 620, 650, 650c
- Carestation 750, 750c
These systems are used in hospitals and surgical settings for delivering anesthesia and ventilatory support to patients of all ages.
The Problem
In VCV mode, the devices affected may fail to ventilate properly. If this happens:
- Alarms and warnings will be triggered (e.g., bellows stop moving, “Unable to Drive Bellows” message, apnea alerts).
- If uncorrected, this can lead to serious injury or death from hypoxia (low oxygen levels).
What to Do
GE HealthCare recommends:
- Do NOT use VCV mode until the device is corrected.
- Use only:
- Pressure Control Ventilation (PCV)
- Pressure Control Ventilation Volume Guarantee (PCV-VG)
- Or manual ventilation
- Perform a Ventilation Screening Test on each system.
- Train your staff and ensure they are aware of the risk and temporary workaround.
- Return the Acknowledgement Form provided in the correction notice.
Questions?
Call GE HealthCare Service at 1-800-437-1171.
If your facility uses these systems, take immediate action to prevent harm. GE HealthCare is currently correcting the devices — no need to remove them from service if alternate modes are used safely.
FDA – Do Not Use Non-piercing Devices to Measure Blood Glucose
FDA Safety Communication On February 21, 2024 the U.S. Food and Drug Administration (FDA) issued a safety communication regarding the use of smartwatches or smart rings that claim to measure blood glucose levels without the need for skin piercing. Its cautionary...
Medos International Recalls CEREBASE DA Guide Sheath
Urgent Class I Recall Notice Medos International Sàrl has issued an urgent recall for the Cerenovus CEREBASE DA Guide Sheath, a neurovascular catheter commonly used in procedures requiring precise navigation and access to blood vessels in the brain. The recall was...
Asensus Surgical Inc Issues Recall for Senhance Surgical System
Class I Recall Notice Asensus Surgical Inc. has announced a recall for its Senhance Surgical System due to malfunctions leading to unintended movement of the robotically-assisted surgical device. Device Use The Senhance Surgical System is a vital tool utilized in...