GE HealthCare Anesthesia Systems Recall

potentially life-threatening risk in Volume Control Ventilation (VCV) mode

GE Carestation Anesthesia Systems

GE HealthCare has issued a high-priority correction for certain Carestation anesthesia systems due to a potentially life-threatening risk when used in Volume Control Ventilation (VCV) mode.

Affected Devices

  • Carestation 620, 650, 650c
  • Carestation 750, 750c

These systems are used in hospitals and surgical settings for delivering anesthesia and ventilatory support to patients of all ages.

The Problem

In VCV mode, the devices affected may fail to ventilate properly. If this happens:

  • Alarms and warnings will be triggered (e.g., bellows stop moving, “Unable to Drive Bellows” message, apnea alerts).
  • If uncorrected, this can lead to serious injury or death from hypoxia (low oxygen levels).

What to Do

GE HealthCare recommends:

  1. Do NOT use VCV mode until the device is corrected.
  2. Use only:
    • Pressure Control Ventilation (PCV)
    • Pressure Control Ventilation Volume Guarantee (PCV-VG)
    • Or manual ventilation
  3. Perform a Ventilation Screening Test on each system.
  4. Train your staff and ensure they are aware of the risk and temporary workaround.
  5. Return the Acknowledgement Form provided in the correction notice.

Questions?

Call GE HealthCare Service at 1-800-437-1171.

If your facility uses these systems, take immediate action to prevent harm. GE HealthCare is currently correcting the devices — no need to remove them from service if alternate modes are used safely.

Electrical Stimulation Devices – ESDs

Devices use controlled electrical pulses Electrical stimulation medical devices are sophisticated tools designed to deliver controlled electrical impulses to specific areas of the body for therapeutic purposes. These devices utilize various forms of electrical...

FDA Proposes Ban of Certain ESDs

Ban would affect devices used to treat aggressive/self-injurious behavior In a landmark move, the U.S. Food and Drug Administration (FDA) has proposed a ban on electrical stimulation devices (ESDs) designed to reduce or stop self-injurious or aggressive behavior. This...

FDA Approves Thoracic Stent Graft System

FDA Approves RelayPro Thoracic Stent Graft System for Thoracic Aortic Aneurysms and Penetrating Atherosclerotic Ulcers Treatment The United States Food and Drug Administration (FDA) on March 7, 2023 approved the RelayPro Thoracic Stent Graft System, a new medical...