GE HealthCare Anesthesia Systems Recall
potentially life-threatening risk in Volume Control Ventilation (VCV) modeGE Carestation Anesthesia Systems
GE HealthCare has issued a high-priority correction for certain Carestation anesthesia systems due to a potentially life-threatening risk when used in Volume Control Ventilation (VCV) mode.
Affected Devices
- Carestation 620, 650, 650c
- Carestation 750, 750c
These systems are used in hospitals and surgical settings for delivering anesthesia and ventilatory support to patients of all ages.
The Problem
In VCV mode, the devices affected may fail to ventilate properly. If this happens:
- Alarms and warnings will be triggered (e.g., bellows stop moving, “Unable to Drive Bellows” message, apnea alerts).
- If uncorrected, this can lead to serious injury or death from hypoxia (low oxygen levels).
What to Do
GE HealthCare recommends:
- Do NOT use VCV mode until the device is corrected.
- Use only:
- Pressure Control Ventilation (PCV)
- Pressure Control Ventilation Volume Guarantee (PCV-VG)
- Or manual ventilation
- Perform a Ventilation Screening Test on each system.
- Train your staff and ensure they are aware of the risk and temporary workaround.
- Return the Acknowledgement Form provided in the correction notice.
Questions?
Call GE HealthCare Service at 1-800-437-1171.
If your facility uses these systems, take immediate action to prevent harm. GE HealthCare is currently correcting the devices — no need to remove them from service if alternate modes are used safely.
A Team-based Approach to an Expedited 510(k)
Getting an expedited 510(k) submission So, you have created a new medical device by following your quality system’s design controls to plan the project, define design inputs, test to verify that outputs meet inputs, and validated your production product. Now what?...
FDA Warning Letter to Olympus Medical Systems
Letter addresses several violations of FDA&C Act On March 15, 2023, the Food and Drug Administration (FDA) issued a Warning Letter to Olympus Medical Systems Corp. The letter addressed several violations of the Federal Food, Drug, and Cosmetic Act (FD&C Act)...
Class I Recall for Intra-Aortic Blood Pumps
Datascope/Getinge IABPs May Malfunction Due to Cable Failure The US Food and Drug Administration (FDA) announced yesterday that Datascope/Getinge is recalling certain CARDIOSAVE Hybrid and Rescue Intra-Aortic Balloon Pumps (IABP) due to a risk of device malfunction....