GE HealthCare Anesthesia Systems Recall

potentially life-threatening risk in Volume Control Ventilation (VCV) mode

GE Carestation Anesthesia Systems

GE HealthCare has issued a high-priority correction for certain Carestation anesthesia systems due to a potentially life-threatening risk when used in Volume Control Ventilation (VCV) mode.

Affected Devices

  • Carestation 620, 650, 650c
  • Carestation 750, 750c

These systems are used in hospitals and surgical settings for delivering anesthesia and ventilatory support to patients of all ages.

The Problem

In VCV mode, the devices affected may fail to ventilate properly. If this happens:

  • Alarms and warnings will be triggered (e.g., bellows stop moving, “Unable to Drive Bellows” message, apnea alerts).
  • If uncorrected, this can lead to serious injury or death from hypoxia (low oxygen levels).

What to Do

GE HealthCare recommends:

  1. Do NOT use VCV mode until the device is corrected.
  2. Use only:
    • Pressure Control Ventilation (PCV)
    • Pressure Control Ventilation Volume Guarantee (PCV-VG)
    • Or manual ventilation
  3. Perform a Ventilation Screening Test on each system.
  4. Train your staff and ensure they are aware of the risk and temporary workaround.
  5. Return the Acknowledgement Form provided in the correction notice.

Questions?

Call GE HealthCare Service at 1-800-437-1171.

If your facility uses these systems, take immediate action to prevent harm. GE HealthCare is currently correcting the devices — no need to remove them from service if alternate modes are used safely.

FDA Issues Letter to Warn of Abbott Trifecta Valve Risks 

Letter outlines risks associated with the heart valve  On February 27, the FDA released a letter to healthcare providers concerning potential risks associated with Abbott Trifecta valves. The Trifecta valve is a type of heart valve that is implanted in patients during...

FDA Announces Recall of GE Nuclear Medicine Systems 

1200-pound Devices Could Fall on Patients On February 15, the FDA issued a statement announcing a recall of certain nuclear medicine systems manufactured by GE Healthcare. The recall involves the 600/800 Series nuclear medicine systems, which are used to diagnose and...

FDA announces recall of LivaNova LifeSPARC System 

The US Food and Drug Administration (FDA) has announced a recall of the Livanova TandemLife LifeSparc System, a medical device used for circulatory support during heart surgery. The recall was initiated due to the potential for the device to experience unintentional...