GE HealthCare Anesthesia Systems Recall
potentially life-threatening risk in Volume Control Ventilation (VCV) modeGE Carestation Anesthesia Systems
GE HealthCare has issued a high-priority correction for certain Carestation anesthesia systems due to a potentially life-threatening risk when used in Volume Control Ventilation (VCV) mode.
Affected Devices
- Carestation 620, 650, 650c
- Carestation 750, 750c
These systems are used in hospitals and surgical settings for delivering anesthesia and ventilatory support to patients of all ages.
The Problem
In VCV mode, the devices affected may fail to ventilate properly. If this happens:
- Alarms and warnings will be triggered (e.g., bellows stop moving, “Unable to Drive Bellows” message, apnea alerts).
- If uncorrected, this can lead to serious injury or death from hypoxia (low oxygen levels).
What to Do
GE HealthCare recommends:
- Do NOT use VCV mode until the device is corrected.
- Use only:
- Pressure Control Ventilation (PCV)
- Pressure Control Ventilation Volume Guarantee (PCV-VG)
- Or manual ventilation
- Perform a Ventilation Screening Test on each system.
- Train your staff and ensure they are aware of the risk and temporary workaround.
- Return the Acknowledgement Form provided in the correction notice.
Questions?
Call GE HealthCare Service at 1-800-437-1171.
If your facility uses these systems, take immediate action to prevent harm. GE HealthCare is currently correcting the devices — no need to remove them from service if alternate modes are used safely.
Hospira Recalls Antibiotic for Presence of Glass Particles
Hospira, Inc., a division of Pfizer, announced a Class I recall of its Vancomycin Injection product due to visible glass particles that were identified in one vial of the drug. The recall involves lot 33045BA of Vancomycin Hydrochloride Injection USP 1.5g single dose...
Three Companies Recall Blood Pressure Medication
All recalled products had unacceptable levels of nitrosamines Three pharmaceutical manufacturers have recalled batches of the blood pressure drug quinapril this year for exceeding safe levels of nitrosamines. Luprin Pharmaceuticals, Aurobindo Pharma USA, and Pfizer...
FDA Announces Online Electronic 510(k) Submissions
The U.S. Food and Drug Administration (FDA) announced the issuance of its final guidance document for its template for electronic 510(k) submissions on September 22, 2022. The electronic Submission Template and Resource (eSTAR) electronic submissions template is a...