AirLife/Vyaire Infant Breathing Systems Recall
Potential for risk of disconnectionClass I Recall
AirLife and manufacturer Vyaire are recalling specific Infant Heated Wire Circuits due to the risk that accessory adapters may disconnect during use, especially once the system reaches operating temperature. This could lead to serious injury or death, including hypoxia, hypercapnia, organ failure, and respiratory collapse.
The FDA has classified this as a Class I recall, the most serious type.
Device Use
These circuits are used to warm breathing gases during mechanical ventilation in neonatal and infant patients, connecting the ventilator to the airway.
Affected Products
- AirLife and Vyaire Infant Heated Wire Circuit AH165 and kits with AH265 chamber
- High-flow dual-limb, dual-heat systems (>4 L/min)
- Full UDI details are listed in the FDA’s Medical Device Recall Database
What You Should Do
- Do NOT use the accessory adapter connections from affected circuit kits.
- Stop using any devices currently using these adapters.
- Devices may still be used if adapter connections are not installed.
- Quarantine affected products, notify relevant staff, and perform regular system checks.
- Activate all ventilator alarms and monitor respiratory parameters closely.
- Ensure backup ventilation/resuscitation equipment is available at bedside.
- Test circuits before use for leaks.
- Return the completed Response Form and notify any customers who received the affected products.
- For urgent needs, call AirLife at 1-800-433-2797 or email: productquality@myairlife.com
This recall is a critical safety action to protect vulnerable infant patients from potential airway disconnection and respiratory failure. Prompt inventory review and staff notification are essential.
Affected Products
| Product Description | Individual UDI | Case UDI |
| AirLife Infant Heated Wire Circuit dual-limb dual-heat, highflow circuit (>4L/min)/AH165 (AirLife) | 10889483595862 | 30889483595866 |
| AirLife Infant Heated Wire Circuit dual-limb dual-heat, highflow circuit (>4L/min)/AH165 (Vyaire) | 10190752145139 | 50190752145137 |
| AirLife Infant Heated Wire Circuit Kit Dual-limb, Dual-heat, high-flow circuit (>4 L/min), Contains AH165 and Circuit, AH290 Chamber/AH265 (AirLife) | 10889483595909 | 30889483595903 |
| AirLife Infant Heated Wire Circuit Kit Dual-limb, Dual-heat, high-flow circuit (>4 L/min), Contains AH165 and Circuit, AH290 Chamber/AH265 (Vyaire) | 10190752145160 | 50190752145168 |
Pharmacal Recalls MG217 Eczema Cream
Pharmacal Recalls MG217 Eczema Cream Due to Staphylococcus aureus ContaminationFDA posts nationwide recall after microbial contamination found in topical skin treatment Pharmacal has voluntarily recalled one lot of MG217 Multi-Symptom Treatment Cream & Skin...
Cefazolin Injection Recalled Due to Cardboard in Sterile IV Solution
Baxter Recalls Cefazolin Injection Sterile IV SolutionCardboard Particulate Matter Found in a Sterile Intravenous Infusion Bag Baxter International has voluntarily recalled one lot of Cefazolin in Dextrose Injection, USP, 2 g/100 mL after a report of cardboard...
FDA Early Alert for Baxter Intravascular Administration Sets
Healthcare facilities are instructed to immediately stop using affected IV solution sets while FDA evaluates the potentially high-risk device issueThe U.S. Food and Drug Administration (FDA) has issued an Early Alert regarding certain Baxter Duo-Vent Solution Sets...