AirLife/Vyaire Infant Breathing Systems Recall

Potential for risk of disconnection

Class I Recall

AirLife and manufacturer Vyaire are recalling specific Infant Heated Wire Circuits due to the risk that accessory adapters may disconnect during use, especially once the system reaches operating temperature. This could lead to serious injury or death, including hypoxia, hypercapnia, organ failure, and respiratory collapse.

The FDA has classified this as a Class I recall, the most serious type.

Device Use

These circuits are used to warm breathing gases during mechanical ventilation in neonatal and infant patients, connecting the ventilator to the airway.

Affected Products

  • AirLife and Vyaire Infant Heated Wire Circuit AH165 and kits with AH265 chamber
  • High-flow dual-limb, dual-heat systems (>4 L/min)
  • Full UDI details are listed in the FDA’s Medical Device Recall Database

What You Should Do

  • Do NOT use the accessory adapter connections from affected circuit kits.
  • Stop using any devices currently using these adapters.
  • Devices may still be used if adapter connections are not installed.
  • Quarantine affected products, notify relevant staff, and perform regular system checks.
  • Activate all ventilator alarms and monitor respiratory parameters closely.
  • Ensure backup ventilation/resuscitation equipment is available at bedside.
  • Test circuits before use for leaks.
  • Return the completed Response Form and notify any customers who received the affected products.
  • For urgent needs, call AirLife at 1-800-433-2797 or email: productquality@myairlife.com

This recall is a critical safety action to protect vulnerable infant patients from potential airway disconnection and respiratory failure. Prompt inventory review and staff notification are essential. 

 

Affected Products

Affected Lot Numbers

Product Description Individual UDI Case UDI
AirLife Infant Heated Wire Circuit dual-limb dual-heat, highflow circuit (>4L/min)/AH165 (AirLife) 10889483595862 30889483595866
AirLife Infant Heated Wire Circuit dual-limb dual-heat, highflow circuit (>4L/min)/AH165 (Vyaire) 10190752145139 50190752145137
AirLife Infant Heated Wire Circuit Kit Dual-limb, Dual-heat, high-flow circuit (>4 L/min), Contains AH165 and Circuit, AH290 Chamber/AH265 (AirLife) 10889483595909 30889483595903
AirLife Infant Heated Wire Circuit Kit Dual-limb, Dual-heat, high-flow circuit (>4 L/min), Contains AH165 and Circuit, AH290 Chamber/AH265 (Vyaire) 10190752145160 50190752145168

FDA Explores Dental Device Safety Concerns

Dental Devices to Treat Sleep Apnea and TMD The FDA is currently evaluating safety concerns related to certain dental devices called fixed (non-removable) palatal expanders, which are used on adults for jaw remodeling or to treat conditions like obstructive sleep...

Update – Carcinoma Inside Capsule of Breast Implants

Squamous Cell Carcinoma and Various Lymphomas Located in the Capsule around Breast Implants On March 22, the US Food and Drug Administration (FDA) issued an update to its March 8 safety communication regarding reports of squamous cell carcinoma (SCC) and lymphomas in...

Carcinoma and Lymphomas Found Near Breast Implants

Squamous Cell Carcinoma and Various Lymphomas Located in the Capsule around Breast Implants On March 8, the US Food and Drug Administration (FDA) issued a safety communication concerning reports of squamous cell carcinoma (SCC) and various types of lymphomas in the...