BD Alaris Pump Module model 8100

Performance can worsen when connected to other perfusion sets

Class I Recall (Correction)

The FDA has issued an early alert to healthcare providers and patients regarding a performance issue affecting BD Alaris Pump Module model 8100, used with specific compatible infusion sets. BD, through its subsidiary CareFusion, has distributed an Urgent Medical Device Recall (Correction) letter outlining concerns about the accuracy of flow rate, bolus delivery, and alarm timing under certain conditions.

What’s the Issue?

The BD Alaris 8100 infusion pump may show worsened performance when used with a subset of compatible pump infusion sets—especially those containing a 0.2-micron in-line filter. The key concerns include:

  • Inaccurate flow rates
  • Delayed alarms
  • Bolus volumes outside expected ranges

These issues could increase the risk of under- or over-infusion in critical care settings.

What Should Users Do?

BD has advised that devices may continue to be used, but specific precautions should be taken to reduce risks. If your facility uses the affected infusion sets, please:

  • Refer to updated performance data available in the BD Alaris User Manuals.
  • Follow all existing warnings and recommendations related to:
    • Flow rate and bolus accuracy
    • Time to alarm
    • Post-occlusion bolus volume (POBV)

For the 0.2-micron in-line filter infusion set specifically:

  • Ensure tubing is not kinked, clamps are open, and drops are flowing in the drip chamber.
  • Avoid infusion rates below 1 mL/h.
  • Do not use pressure mode settings above 50 mmHg for rates under 10 mL/h.

Facility Action Items

  • Post the BD Customer Notification near all affected devices.
  • Share the notice with all staff responsible for infusion pump use.
  • If issues are encountered, report them to BD at: productcomplaints@bd.com

Stay Updated

The FDA continues to evaluate this issue as potentially high-risk and will update the public with any new findings. For the most current information, visit the FDA’s Medical Device Recalls page.

If you need help identifying affected equipment or implementing corrective actions, don’t hesitate to reach out to your BD representative or internal clinical engineering team.

Early Alert: Infusion Set Performance Issue from BD

The FDA is aware that BD and their subsidiary CareFusion have issued an “Urgent Medical Device Recall (Correction)” letter notifying affected customers of worse performance, under certain use cases, for the BD Alaris Pump Module model 8100 (pump module) when used with a subset of compatible pump infusion sets.

What to Do

Refer to the updated set performance data as compared to the existing set performance data found in User Manuals. Follow applicable warnings and recommendations from the User Manual to mitigate risks associated with these infusion parameters. Consider using additional mitigations with the 0.2-micron in-line filter infusion set.

On July 8, BD sent all affected customers a letter recommending the following actions:

  • Devices can continue to be used as intended.
  • The User Manuals currently provide applicable warnings and recommendations for clinical users to mitigate the risks associated with conditions that shift flow rate accuracy, time to alarm, and bolus accuracy performance.
    • Please refer to the recommendations and warnings from the BD Alaris User Manual regarding rate and bolus accuracy, time to alarm, and post-occlusion bolus volume (POBV).
  • The following actions should be considered for the 0.2-micron in-line filter infusion set:
    • Ensure all clamps are open; there are no kinks or collapses in the tubing outside of the pump; drops are flowing in the drip chamber.
    • Avoid infusion flow rates below 1 mL/h.
    • Avoid using Selectable pressure mode with a setting above 50 mmHg for rates less than 10 mL/h.
  • Post this Customer Notification on or near the affected products. If you experience any issues, contact BD at productcomplaints@bd.com
  • Circulate this notice within your facility network to ensure that all concerned personnel are made aware of this issue.

Check the FDA recall web page for updates. The FDA is currently reviewing information about this potentially high-risk device issue and will keep the public informed as significant new information becomes available.

BD Alaris pump module

A BD Aleris pump module
Source: FDA

Now the Salmonella in Moringa Is Drug-Resistant

Now the Salmonella in Moringa Is Drug-Resistant. That Changes the Stakes. An MDP Consumer and Industry Brief Salmonella contamination is serious in any consumer product. But one of the most concerning developments in the latest moringa-related outbreak investigation...

Four Moringa-Linked Salmonella Events in Four Months

Four Moringa-Linked Salmonella Events in Four Months: The Timeline That Should Not Exist An MDP Consumer and Industry Brief A single recall can be dismissed as a one-time failure. Two recalls in the same ingredient category raise questions. But when moringa-based...

Rosabella Moringa Capsules Potential Salmonella Contamination

Rosabella Moringa Capsules Recalled Over Possible Salmonella Contamination Editor's Note: This February 13 recall is the latest moringa-root product recall, following the Sam's Super Greens recall in November, the Live it Up Super Greens recall on January 15, and the...

Affected Products

Product Name 

Catalog/Model (Ref) No.

Software Version

UDI-DI

BD Alaris Pump Module Model 8100 

8100

All software versions

10885403222054
10885403517723
10885403810015
10885403810039
10885403810046

BD Alaris System with Guardrails Suite MX (PCU)

8015

All software versions up to v12.5

10885403801549
10885403801532
10885403801518
10885403519284
10885403519215
10885403517921
10885403516023
10885403515231
10885403519291

BD Alaris Pump Infusion Set, Back Check Valve, 3 SmartSite Y-sites

10013186

NA 

07613203021173

BD Alaris Pump Infusion Set, 2 Back Check Valves, 3 SmartSite Y-sites

2452-0007

NA 

10885403219870

BD Alaris Pump Infusion Set

2204-0007

NA 

10885403199363

BD Alaris Pump Infusion Set, Back Check Valve, SmartSite Y-site

24001-0007

NA 

10885403238666

BD Alaris Pump Infusion Burette Set, 0.2 Micron ilter, Smallbore Tubing, SmartSite Port (Burette), SmartSite Y-site

10015012

NA 

10885403233951

BD Alaris Pump Infusion Set, 0.2 Micron Filter, Back Check Valve, 3 SmartSite Y-sites

2432-0007

NA 

10885403232329

BD Alaris Pump Infusion Set, 0.2 Micron Filter, Back Check Valve, SmartSite Y-site

2434-0007

NA 

07613203019682

BD Alaris Pump Infusion Set, 1.2 Micron Filter

2202-0007

NA 

10885403274039

BD Alaris Pump Infusion Set, Back Check Valve, 5 SmartSite Y-sites

11426965

NA 

10885403232558

BD Alaris Pump Infusion Set, 15 Micron Filter, Back Check Valve, 3 SmartSite Y-sites

10561554

NA 

10885403232565

BD Alaris Pump Infusion Set, 15 Micron Filter

10863358

NA 

10885403232466

BD Alaris Pump Infusion Set, 2 Back Check Valves, 3 SmartSite Y-sites

11171447

NA 

07613203021234

BD Alaris Pump Infusion Set, 0.2 Micron Filter, Back Check Valve, PE Lined Tubing, 2 SmartSite Y-sites

11532269

NA 

10885403232343

BD Alaris Pump Infusion Set, 0.2 Micron Filter, PE Lined Tubing, SmartSite Y-site

10010454

NA 

07613203015806

BD Alaris Pump Infusion Set, SmartSite Bag Access Non-Vented, 0.2 Micron Filter, PE Lined Tubing, SmartSite Y-site

2465-0007

NA 

10885403221941

BD Alaris Pump Infusion Set, 1.2 Micron Filter, PE Lined Tubing, SmartSite Y-site

10010453

NA 

07613203021135