Cranial Drill Urgent Recall Alert

Serious safety risks

Urgent Class I Medical Device Recall

Integra LifeSciences has issued an urgent recall of specific Codman Disposable Perforators and Craniotomy Kits due to a serious manufacturing defect that may lead to device disassembly during neurosurgical procedures. The FDA has classified this as a Class I recall, the most serious type, due to the risk of severe injury or death.

Affected Products

  • Codman Disposable Perforator 14 mm – UDI: 10381780513599
  • Codman Craniotomy Kit (includes Perforator, Cranio-blade, Wire Pass Drill) – UDI: 10381780513629

Lot/Serial Numbers: Full List of Affected Lots

Reason for Recall

A faulty ultrasonic weld (known as a “proud weld”) on the device’s outer sleeve may cause the perforator to disassemble unexpectedly during use. In some cases, the device fails to disengage properly, putting patients at serious risk for:

  • Brain tissue injury
  • Bleeding or dural damage
  • Extended surgery time
  • Irreversible harm or even death

To date, 10 serious injuries have been reported—including devices lodged in the skull and fragments difficult to remove. No fatalities have been reported.

Recommended Actions for Users and Facilities

  • Immediately stop using and quarantine all affected products.
  • Review inventory against the provided lot numbers.
  • Notify surgical and supply staff of the recall.
  • Contact Integra for a Return Material Authorization (RMA) and return the product for credit.

Device Function

Codman Disposable Perforators are single-use neurosurgical tools designed to drill precise openings into the skull during craniotomy. They are engineered to disengage automatically after drilling to prevent over-penetration—but this defect undermines that safety feature.

Contact

For questions or to report issues, call Integra LifeSciences at 1-800-654-2873 or email FCA3@integralife.com.

Check the FDA recall web page for updates. The FDA is currently reviewing information about this potentially high-risk device issue and will keep the public informed as significant new information becomes available.

Urgent Recall Notice for ARROW QuickFlash Catheterization Kits 

Urgent Recall Notice for ARROW QuickFlash Catheterization Kits  In February 2024, Teleflex and its subsidiary Arrow International issued an urgent recall notice for the ARROW QuickFlash Radial Artery and Radial Artery/Arterial Line Catheterization Kits.   Today, a...

Plastic Syringe Safety: FDA’s Ongoing Evaluation

Today, the U.S. Food and Drug Administration (FDA) provided a crucial update regarding the quality and performance issues associated with plastic syringes made in China. This announcement follows the agency's previous safety communication issued on November 30, 2023,...

Original FDA Safety Warning for China-made Syringes

In an announcement dated November 30, 2023, the U.S. Food and Drug Administration (FDA) issued a notice regarding the evaluation of potential device failures associated with plastic syringes manufactured in China. The FDA is actively collecting and analyzing data to...