Cranial Drill Urgent Recall Alert

Serious safety risks

Urgent Class I Medical Device Recall

Integra LifeSciences has issued an urgent recall of specific Codman Disposable Perforators and Craniotomy Kits due to a serious manufacturing defect that may lead to device disassembly during neurosurgical procedures. The FDA has classified this as a Class I recall, the most serious type, due to the risk of severe injury or death.

Affected Products

  • Codman Disposable Perforator 14 mm – UDI: 10381780513599
  • Codman Craniotomy Kit (includes Perforator, Cranio-blade, Wire Pass Drill) – UDI: 10381780513629

Lot/Serial Numbers: Full List of Affected Lots

Reason for Recall

A faulty ultrasonic weld (known as a “proud weld”) on the device’s outer sleeve may cause the perforator to disassemble unexpectedly during use. In some cases, the device fails to disengage properly, putting patients at serious risk for:

  • Brain tissue injury
  • Bleeding or dural damage
  • Extended surgery time
  • Irreversible harm or even death

To date, 10 serious injuries have been reported—including devices lodged in the skull and fragments difficult to remove. No fatalities have been reported.

Recommended Actions for Users and Facilities

  • Immediately stop using and quarantine all affected products.
  • Review inventory against the provided lot numbers.
  • Notify surgical and supply staff of the recall.
  • Contact Integra for a Return Material Authorization (RMA) and return the product for credit.

Device Function

Codman Disposable Perforators are single-use neurosurgical tools designed to drill precise openings into the skull during craniotomy. They are engineered to disengage automatically after drilling to prevent over-penetration—but this defect undermines that safety feature.

Contact

For questions or to report issues, call Integra LifeSciences at 1-800-654-2873 or email FCA3@integralife.com.

Check the FDA recall web page for updates. The FDA is currently reviewing information about this potentially high-risk device issue and will keep the public informed as significant new information becomes available.

FDA Issues Letter to Warn of Abbott Trifecta Valve Risks 

Letter outlines risks associated with the heart valve  On February 27, the FDA released a letter to healthcare providers concerning potential risks associated with Abbott Trifecta valves. The Trifecta valve is a type of heart valve that is implanted in patients during...

FDA Announces Recall of GE Nuclear Medicine Systems 

1200-pound Devices Could Fall on Patients On February 15, the FDA issued a statement announcing a recall of certain nuclear medicine systems manufactured by GE Healthcare. The recall involves the 600/800 Series nuclear medicine systems, which are used to diagnose and...

FDA announces recall of LivaNova LifeSPARC System 

The US Food and Drug Administration (FDA) has announced a recall of the Livanova TandemLife LifeSparc System, a medical device used for circulatory support during heart surgery. The recall was initiated due to the potential for the device to experience unintentional...