Cranial Drill Urgent Recall Alert

Serious safety risks

Urgent Class I Medical Device Recall

Integra LifeSciences has issued an urgent recall of specific Codman Disposable Perforators and Craniotomy Kits due to a serious manufacturing defect that may lead to device disassembly during neurosurgical procedures. The FDA has classified this as a Class I recall, the most serious type, due to the risk of severe injury or death.

Affected Products

  • Codman Disposable Perforator 14 mm – UDI: 10381780513599
  • Codman Craniotomy Kit (includes Perforator, Cranio-blade, Wire Pass Drill) – UDI: 10381780513629

Lot/Serial Numbers: Full List of Affected Lots

Reason for Recall

A faulty ultrasonic weld (known as a “proud weld”) on the device’s outer sleeve may cause the perforator to disassemble unexpectedly during use. In some cases, the device fails to disengage properly, putting patients at serious risk for:

  • Brain tissue injury
  • Bleeding or dural damage
  • Extended surgery time
  • Irreversible harm or even death

To date, 10 serious injuries have been reported—including devices lodged in the skull and fragments difficult to remove. No fatalities have been reported.

Recommended Actions for Users and Facilities

  • Immediately stop using and quarantine all affected products.
  • Review inventory against the provided lot numbers.
  • Notify surgical and supply staff of the recall.
  • Contact Integra for a Return Material Authorization (RMA) and return the product for credit.

Device Function

Codman Disposable Perforators are single-use neurosurgical tools designed to drill precise openings into the skull during craniotomy. They are engineered to disengage automatically after drilling to prevent over-penetration—but this defect undermines that safety feature.

Contact

For questions or to report issues, call Integra LifeSciences at 1-800-654-2873 or email FCA3@integralife.com.

Check the FDA recall web page for updates. The FDA is currently reviewing information about this potentially high-risk device issue and will keep the public informed as significant new information becomes available.

Global Medical Device Reporting and Vigilance

International Medical Device Reporting Medical Device Reporting and Vigilance in the EU and Globally  Timely and transparent reporting of medical device incidents is critical to protecting public health and ensuring continued confidence in healthcare technologies....

Device Recall Alert: Edwards Lifesciences Arterial Cannula

Edwards Lifesciences Arterial CannulaSource: FDAEdwards Lifesciences Recalls Arterial Cannulae Over Risk of Exposed Wire Edwards Lifesciences has issued a Class I recall of certain arterial cannula devices due to the risk of wire exposure near the cannula tip, which...

Device Recall: ZOLL AutoPulse NXT Resuscitation System

Urgent Class I Medical Device Recall ZOLL Circulation, Inc. has issued a Class I recall of its AutoPulse NXT Resuscitation System (Model 200, Software Version 2.0.1) due to a system error that may cause the device to stop chest compressions or deliver compressions...