FDA Alert: Abiomed Impella Blood Pump Urgent Recall
Controller may fail to detect the Impella deviceAbiomed Impella blood pump controllers
Controller May Fail to Detect Device
Abiomed has issued an urgent safety notice for its Automated Impella Controller (AIC) due to a potentially serious issue where the system may fail to detect the Impella blood pump when connected.
This malfunction could result in inadequate cardiac support, which is especially dangerous for patients in cardiogenic shock.
Affected Products, Product Codes, UDI-DIs, and Serial Numbers (click on the link to download a list.)
Key Details
- Issue: The controller may not recognize the connected pump during startup or transfer between consoles.
- No visual alarm appears on screen to indicate a problem.
- Reported Cases: 3 patient deaths have been linked to this issue as of June 13; no serious injuries reported.
- Impacted Devices: All AICs across the Abiomed Impella pump line.
Recommended Actions
- Always have a backup AIC available.
- If the pump isn’t detected within 20 seconds:
- Switch back to the previous console.
- Reboot the non-responsive console.
- Start the case on a different AIC if needed.
- Distribute this alert to all relevant clinical staff and facilities.
- Forward notice to any facility that received affected products.
Why It Matters
Failure to detect the pump could lead to delayed or absent hemodynamic support, putting critically ill patients at high risk.
For questions or to report adverse events, contact Abiomed at (978) 646-1400.
The FDA is actively reviewing this high-risk issue and will provide updates here as more information becomes available.
Read more about Mechanical Circulatory Support (MCS) devices
QMSR Management Responsibility
QMSR Management Responsibility The QMSR Management Responsibility requirements bring a more structured, documented, and proactive approach compared to QSR §820.20. While many concepts remain similar, organizations must focus on: - Stronger documentation of leadership...
QMSR Quality Manual
Why a Quality Manual Matters in the QMSR Transition U.S. Manufacturers Adopting ISO 13485 Will Need a Quality Manual While QSR does not mandate a Quality Manual, the QMSR ISO 13485 compliance does. This means manufacturers aligning with QMSR who do not have a quality...
QMSR Design and Development Transfer and Changes
ISO 13485:2016 Design and Development Clause 7.3 - Design and DevelopmentClause 7.3.1 - GeneralClause 7.3.2 - Design and development planningClause 7.3.3 - Design and development inputsClause 7.3.4 - Design and development outputsClause 7.3.5 - Design and development...