FDA Alert: Abiomed Impella Blood Pump Urgent Recall

Controller may fail to detect the Impella device

Abiomed Impella blood pump controllers

Controller May Fail to Detect Device

Abiomed has issued an urgent safety notice for its Automated Impella Controller (AIC) due to a potentially serious issue where the system may fail to detect the Impella blood pump when connected.

This malfunction could result in inadequate cardiac support, which is especially dangerous for patients in cardiogenic shock.

Affected Products, Product Codes, UDI-DIs, and Serial Numbers (click on the link to download a list.)

Key Details

  • Issue: The controller may not recognize the connected pump during startup or transfer between consoles.
  • No visual alarm appears on screen to indicate a problem.
  • Reported Cases: 3 patient deaths have been linked to this issue as of June 13; no serious injuries reported.
  • Impacted Devices: All AICs across the Abiomed Impella pump line.

Recommended Actions

  • Always have a backup AIC available.
  • If the pump isn’t detected within 20 seconds:
    • Switch back to the previous console.
    • Reboot the non-responsive console.
    • Start the case on a different AIC if needed.
  • Distribute this alert to all relevant clinical staff and facilities.
  • Forward notice to any facility that received affected products.

Why It Matters

Failure to detect the pump could lead to delayed or absent hemodynamic support, putting critically ill patients at high risk.

For questions or to report adverse events, contact Abiomed at (978) 646-1400.

The FDA is actively reviewing this high-risk issue and will provide updates here as more information becomes available.

Read more about Mechanical Circulatory Support (MCS) devices

Mechanical Circulatory Support (MCS) Devices

Overview of MCS Devices Mechanical Circulatory Support (MCS) devices have transformed the landscape of advanced heart failure treatment. From temporary stabilization in cardiogenic shock to long-term support as destination therapy, these technologies bridge critical...

Recall Alert: AirLife/Vyaire Infant Breathing Systems

Class I Recall AirLife and manufacturer Vyaire are recalling specific Infant Heated Wire Circuits due to the risk that accessory adapters may disconnect during use, especially once the system reaches operating temperature. This could lead to serious injury or death,...

Avanos Recalls Ballard Closed Suction Systems

May result in non-sterile devices Avanos Medical, Inc. is recalling its Ballard Closed Suction Systems after identifying a failure in the sterilization process, which may result in non-sterile devices. Use of these products could lead to serious health risks,...