GE HealthCare Anesthesia Systems Recall

potentially life-threatening risk in Volume Control Ventilation (VCV) mode

GE Carestation Anesthesia Systems

GE HealthCare has issued a high-priority correction for certain Carestation anesthesia systems due to a potentially life-threatening risk when used in Volume Control Ventilation (VCV) mode.

Affected Devices

  • Carestation 620, 650, 650c
  • Carestation 750, 750c

These systems are used in hospitals and surgical settings for delivering anesthesia and ventilatory support to patients of all ages.

The Problem

In VCV mode, the devices affected may fail to ventilate properly. If this happens:

  • Alarms and warnings will be triggered (e.g., bellows stop moving, “Unable to Drive Bellows” message, apnea alerts).
  • If uncorrected, this can lead to serious injury or death from hypoxia (low oxygen levels).

What to Do

GE HealthCare recommends:

  1. Do NOT use VCV mode until the device is corrected.
  2. Use only:
    • Pressure Control Ventilation (PCV)
    • Pressure Control Ventilation Volume Guarantee (PCV-VG)
    • Or manual ventilation
  3. Perform a Ventilation Screening Test on each system.
  4. Train your staff and ensure they are aware of the risk and temporary workaround.
  5. Return the Acknowledgement Form provided in the correction notice.

Questions?

Call GE HealthCare Service at 1-800-437-1171.

If your facility uses these systems, take immediate action to prevent harm. GE HealthCare is currently correcting the devices — no need to remove them from service if alternate modes are used safely.

April 2026 Update for FDA QMSR Inspection Results

April 2026 Update for FDA QMSR Inspection ResultsConfirming the Shift to System-Level, Risk-Based Inspection When FDA’s Quality Management System Regulation (QMSR) took effect on February 2, 2026, the regulatory framework became formally aligned with ISO 13485. What...

Merit Medical Dialysis Catheter Recall – Risk of Serious Injury

Merit Medical Removes 16F Dual-Valved Splittable Sheath Introducer Design Defect May Lead to Hemorrhage, Embolization, and Loss of Vascular Access The U.S. Food and Drug Administration (FDA) has identified a Class I recall involving Merit Medical’s 16F Dual-Valved...

Early Alert: Cook Medical Sizing Catheters

Cook Medical Sizing Catheters Removed Due to Risk of Device Breakage The U.S. Food and Drug Administration (FDA) has issued an early alert regarding a potentially serious device issue involving certain sizing catheters manufactured by Cook Medical. Cook Medical has...