GE HealthCare Anesthesia Systems Recall
potentially life-threatening risk in Volume Control Ventilation (VCV) modeGE Carestation Anesthesia Systems
GE HealthCare has issued a high-priority correction for certain Carestation anesthesia systems due to a potentially life-threatening risk when used in Volume Control Ventilation (VCV) mode.
Affected Devices
- Carestation 620, 650, 650c
- Carestation 750, 750c
These systems are used in hospitals and surgical settings for delivering anesthesia and ventilatory support to patients of all ages.
The Problem
In VCV mode, the devices affected may fail to ventilate properly. If this happens:
- Alarms and warnings will be triggered (e.g., bellows stop moving, “Unable to Drive Bellows” message, apnea alerts).
- If uncorrected, this can lead to serious injury or death from hypoxia (low oxygen levels).
What to Do
GE HealthCare recommends:
- Do NOT use VCV mode until the device is corrected.
- Use only:
- Pressure Control Ventilation (PCV)
- Pressure Control Ventilation Volume Guarantee (PCV-VG)
- Or manual ventilation
- Perform a Ventilation Screening Test on each system.
- Train your staff and ensure they are aware of the risk and temporary workaround.
- Return the Acknowledgement Form provided in the correction notice.
Questions?
Call GE HealthCare Service at 1-800-437-1171.
If your facility uses these systems, take immediate action to prevent harm. GE HealthCare is currently correcting the devices — no need to remove them from service if alternate modes are used safely.
Medline Recalls Angiographic RA Control Syringes
Angiographic Syringe Recall: Medline Removes RA Control Syringes Due to Risk of Disconnection and Air Embolism The U.S. Food and Drug Administration (FDA) has announced that Medline Industries has issued an Urgent Medical Device Recall involving certain Namic...
FDA Class I Recall for Abiomed Impella Gen 1 Purge Cassettes
Medical Device Alert: Abiomed Impella Purge Cassette Recall (Class I) Abiomed has initiated a Class I recall, the most serious classification issued by the U.S. Food and Drug Administration, for certain Impella purge cassettes and associated pump sets due to a...
Class I Correction for Abiomed Impella Inaccurate Flow Readings
Abiomed Impella RP: FDA Class I Correction for Inaccurate Flow Readings Abiomed has issued an Urgent Medical Device Correction for its Impella RP with SmartAssist and Impella RP Flex with SmartAssist systems due to potential inaccuracies in displayed pump performance...