AirLife/Vyaire Infant Breathing Systems Recall

Potential for risk of disconnection

Class I Recall

AirLife and manufacturer Vyaire are recalling specific Infant Heated Wire Circuits due to the risk that accessory adapters may disconnect during use, especially once the system reaches operating temperature. This could lead to serious injury or death, including hypoxia, hypercapnia, organ failure, and respiratory collapse.

The FDA has classified this as a Class I recall, the most serious type.

Device Use

These circuits are used to warm breathing gases during mechanical ventilation in neonatal and infant patients, connecting the ventilator to the airway.

Affected Products

  • AirLife and Vyaire Infant Heated Wire Circuit AH165 and kits with AH265 chamber
  • High-flow dual-limb, dual-heat systems (>4 L/min)
  • Full UDI details are listed in the FDA’s Medical Device Recall Database

What You Should Do

  • Do NOT use the accessory adapter connections from affected circuit kits.
  • Stop using any devices currently using these adapters.
  • Devices may still be used if adapter connections are not installed.
  • Quarantine affected products, notify relevant staff, and perform regular system checks.
  • Activate all ventilator alarms and monitor respiratory parameters closely.
  • Ensure backup ventilation/resuscitation equipment is available at bedside.
  • Test circuits before use for leaks.
  • Return the completed Response Form and notify any customers who received the affected products.
  • For urgent needs, call AirLife at 1-800-433-2797 or email: productquality@myairlife.com

This recall is a critical safety action to protect vulnerable infant patients from potential airway disconnection and respiratory failure. Prompt inventory review and staff notification are essential. 

 

Affected Products

Affected Lot Numbers

Product Description Individual UDI Case UDI
AirLife Infant Heated Wire Circuit dual-limb dual-heat, highflow circuit (>4L/min)/AH165 (AirLife) 10889483595862 30889483595866
AirLife Infant Heated Wire Circuit dual-limb dual-heat, highflow circuit (>4L/min)/AH165 (Vyaire) 10190752145139 50190752145137
AirLife Infant Heated Wire Circuit Kit Dual-limb, Dual-heat, high-flow circuit (>4 L/min), Contains AH165 and Circuit, AH290 Chamber/AH265 (AirLife) 10889483595909 30889483595903
AirLife Infant Heated Wire Circuit Kit Dual-limb, Dual-heat, high-flow circuit (>4 L/min), Contains AH165 and Circuit, AH290 Chamber/AH265 (Vyaire) 10190752145160 50190752145168

Class I Correction for Homecare Bed Fire and Entrapment Risks

Medline Homecare Beds: FDA Class I Correction for Fire and Entrapment Risks Medline Industries has issued a critical safety correction involving multiple models of homecare beds following reports of overheating components, electrical hazards, and patient entrapment...

When a Small Process Failure Becomes a Patient Safety Event

What the Erbe Cryoprobe Alert Teaches Us About Modern Device Quality The FDA’s recent Early Alert regarding flexible cryoprobes manufactured by Erbe USA is a textbook example of a fundamental truth in medical device quality. That small manufacturing deviations can...

FDA Flags High-Risk Issue with Erbe Flexible Cryoprobes

FDA Flags High-Risk Issue with Erbe Flexible Cryoprobes The FDA’s Center for Devices and Radiological Health (CDRH) has issued an Early Alert regarding a potentially high-risk issue involving certain flexible cryoprobes manufactured by Erbe USA. Issue Overview Erbe...