HemosIL-AcuStar

HemosIL AcuStar HIT-IgG(PF4-H). Source: fda.gov

Performance problems could produce false-negative results, potentially delaying diagnosis of a dangerous reaction to heparin.

The U.S. Food and Drug Administration (FDA) has identified a recall of certain lots of the HemosIL AcuStar HIT-IgG(PF4-H) diagnostic test, manufactured by Instrumentation Laboratories (Werfen), as the most serious type of medical device recall.

The recall involves a performance problem that could cause the test to incorrectly report negative results in patients being evaluated for heparin-induced thrombocytopenia (HIT), a potentially life-threatening reaction to the blood-thinning medication heparin.

Why the Recall Matters

Heparin is commonly used in hospitals to prevent or treat blood clots. In some patients, however, it can trigger an immune reaction that reduces platelet counts and increases the risk of dangerous blood clots.

The HemosIL AcuStar HIT-IgG(PF4-H) test helps clinicians evaluate patients for this condition by detecting antibodies associated with HIT.

According to the FDA, customer complaints identified problems with test calibration and inaccurate results. Werfen subsequently confirmed that certain lots could produce results with a negative bias of 20% to 35%, meaning the test could report antibody levels lower than their actual values.

More concerning, the problem may not be reliably detected through routine calibration and quality control testing.

As a result, some patient samples could be incorrectly reported below the test’s cutoff of 1.00 U/mL and interpreted as negative for HIT antibodies.

A false-negative result could delay recognition of HIT and influence treatment decisions. Continued heparin exposure in a patient with HIT could increase the risk of serious thrombotic complications.

Products Affected

The recall affects four lots of HemosIL AcuStar HIT-IgG(PF4-H):

Part number Affected lot Expiration date
0009800127 B40005 January 16, 2027
0009800127 B38062 November 1, 2026
0009802028 B40006 January 16, 2027
0009802028 B38063 November 1, 2026

Actions for Healthcare Facilities

Werfen has instructed affected customers to identify and destroy remaining products from the recalled lots, document the destruction, and arrange alternative HIT antibody testing methods. The company has also warned that a supply interruption is expected. Werfen’s investigation into the root cause remains ongoing.

Laboratories and treating clinicians should determine whether previously reported patient results require review, follow-up, or repeat testing based on clinical judgment and applicable procedures.

As of September 3, 2026, the company had reported no serious injuries or deaths associated with the issue.

Regulatory Intelligence Perspective

This recall illustrates an important challenge involving diagnostic devices, that a test can produce an apparently valid result even when its analytical performance is compromised. The reported negative bias is particularly concerning because routine calibration and quality control procedures may not reliably identify the problem.

For diagnostic manufacturers and clinical laboratories, the recall underscores the importance of investigating performance complaints, evaluating potential effects on previously reported patient results, and assessing whether existing quality controls are capable of detecting clinically significant failures.

The case also demonstrates why postmarket surveillance and effective corrective action processes are essential throughout a diagnostic device’s commercial life.

Additional information

U.S. customers may contact Werfen Client Services at 1-800-955-9525. FDA recall information is available through the agency’s Medical Device Recalls and Early Alerts website.

FDA Issues Early Alert for Resmed Astral Ventilators

Resmed VentilatorPatients should continue using devices unless instructed otherwise by their healthcare provider The U.S. Food and Drug Administration (FDA) has issued an Early Alert regarding a potentially high-risk issue affecting certain Resmed Astral 100 and...

One Contaminated Alcohol Pad, Multiple Recalls

A Lesson in Medical Device Supply Chain Quality At first glance, Cardinal Health's recent recall of Webcol™ Alcohol Prep Pads appears straightforward. A commonly used healthcare product was found to be contaminated with Paenibacillus phoenicis, creating a potential...

Early Alert: Windstone Medical Recall for Convenience Kits

FDA Warns of Convenience Kits Containing Recalled Cardinal Health Alcohol Pads The FDA has issued an Early Alert regarding several medical convenience kits distributed by Windstone Medical Packaging, Inc. that contain alcohol prep pads previously recalled by Cardinal...

Webcol Alcohol Prep Pads Recall Due to Microbial Contamination

Cardinal Health Recalls Alcohol Prep Pads Over Infection Risk Concerns Cardinal Health has issued a voluntary nationwide recall of select lots of its Webcol™ Large Alcohol Prep Pads (70% isopropyl alcohol) after testing identified microbial contamination with...

FDA Issues Warning Letter for Modifications to Infant Products

FDA Issues Warning Letter to Happiest Baby Over SNOO Smart Sleeper Modifications and Quality System Deficiencies The U.S. Food and Drug Administration has issued a Warning Letter to Happiest Baby, Inc., manufacturer of the SNOO Smart Sleeper, citing unapproved device...

FDA Expands Draeger Atlan Anesthesia Workstation Correction

More anesthesia machines added to correction for potential ventilation failure The U.S. Food and Drug Administration (FDA) has designated a recent expansion of Draeger, Inc.'s Atlan anesthesia workstation correction as a Class I recall, the agency's most serious...

Warning Letter to Zoll Medical for Major QMS Violations

Warning Letter to Zoll Medical for Major QMS Violations The U.S. Food and Drug Administration has issued a Warning Letter to ZOLL Medical Corporation citing significant Quality System Regulation (QSR) violations involving corrective and preventive action (CAPA),...

Abiomed Issues Correction for Automated Impella Controllers 

Hardware Updates Required to Address Potential Loss of Hemodynamic Support  Abiomed has issued a correction for certain Automated Impella Controllers (AICs) used with the Impella heart pump system after identifying hardware-related issues that could lead to controller...

Infant Formula Recalled after Reports of Botulism

Three Infants Hospitalized; Nationwide Recall Initiated  Nara Organics has voluntarily recalled all lots of its powdered infant formula currently on the U.S. market following reports of three cases of infant botulism in infants who reportedly consumed the product. ...

FDA Clears First OTC Continuous Glucose Monitor for Children

Regulatory Snapshot Device: Stelo® Glucose Biosensor System (Dexcom, Inc.)Indication: Continuous glucose monitoring for individuals ages 2 years and older who do not use insulin, including children with diabetes managed through oral medications and individuals seeking...

Medical Devices
Medical Device Recalls
Medical Device Warning Letters