FDA Alert: Abiomed Impella Blood Pump Urgent Recall

Controller may fail to detect the Impella device

Abiomed Impella blood pump controllers

Controller May Fail to Detect Device

Abiomed has issued an urgent safety notice for its Automated Impella Controller (AIC) due to a potentially serious issue where the system may fail to detect the Impella blood pump when connected.

This malfunction could result in inadequate cardiac support, which is especially dangerous for patients in cardiogenic shock.

Affected Products, Product Codes, UDI-DIs, and Serial Numbers (click on the link to download a list.)

Key Details

  • Issue: The controller may not recognize the connected pump during startup or transfer between consoles.
  • No visual alarm appears on screen to indicate a problem.
  • Reported Cases: 3 patient deaths have been linked to this issue as of June 13; no serious injuries reported.
  • Impacted Devices: All AICs across the Abiomed Impella pump line.

Recommended Actions

  • Always have a backup AIC available.
  • If the pump isn’t detected within 20 seconds:
    • Switch back to the previous console.
    • Reboot the non-responsive console.
    • Start the case on a different AIC if needed.
  • Distribute this alert to all relevant clinical staff and facilities.
  • Forward notice to any facility that received affected products.

Why It Matters

Failure to detect the pump could lead to delayed or absent hemodynamic support, putting critically ill patients at high risk.

For questions or to report adverse events, contact Abiomed at (978) 646-1400.

The FDA is actively reviewing this high-risk issue and will provide updates here as more information becomes available.

Read more about Mechanical Circulatory Support (MCS) devices

April 2026 Update for FDA QMSR Inspection Results

April 2026 Update for FDA QMSR Inspection ResultsConfirming the Shift to System-Level, Risk-Based Inspection When FDA’s Quality Management System Regulation (QMSR) took effect on February 2, 2026, the regulatory framework became formally aligned with ISO 13485. What...

Merit Medical Dialysis Catheter Recall – Risk of Serious Injury

Merit Medical Removes 16F Dual-Valved Splittable Sheath Introducer Design Defect May Lead to Hemorrhage, Embolization, and Loss of Vascular Access The U.S. Food and Drug Administration (FDA) has identified a Class I recall involving Merit Medical’s 16F Dual-Valved...

Early Alert: Cook Medical Sizing Catheters

Cook Medical Sizing Catheters Removed Due to Risk of Device Breakage The U.S. Food and Drug Administration (FDA) has issued an early alert regarding a potentially serious device issue involving certain sizing catheters manufactured by Cook Medical. Cook Medical has...